76 . RamanStation 400 Series Getting Started Guide
If you exit Spectrum software the instrument settings will be remembered when you restart
Spectrum, providing that they have been used to collect data. If you want to return to the
default instrument settings, click
Restore Defaults
on the Setup Instrument Basic tab. If you
wish these parameters to be stored as a new set of instrument settings, select Load and
Save on the Setup Instrument Basic tab.
If you would like to export the current instrument settings as a *.REX for use with AssureID,
click
Export
.
NOTE: For information on how to use AssureID with a triggered fiber optic probe, refer to
the AssureID on-screen help.
NOTE: For information on how use the Setup Sample Area XYZ Stage tab to set up Markers,
Cell Markers, Maps and Line Scans, and to collect a Visible Image Survey, refer to the
Spectrum on-screen help.
SOP-5: Sample Positioning using the Motorized XYZ
Stage
You can use Monitor to help position your sample for optimum spectral collection efficiency.
Sample positioning and alignment in Monitor mode allows you to position and focus on a
specific area and to collect spectral data from that location. It can also be used to enhance
the signal-to-noise ratio of spectral data. The data displayed on screen during monitoring
cannot be saved. To save data you must collect spectra (SOP-4).
NOTE: To see the Live display, the
Show live display
option must be selected on the Setup
Instrument Data Collection tab.
1.
Ensure the RamanStation 400 is switched on.
If not, follow SOP-1.
2.
Load the appropriate sample holder onto the motorized XYZ stage and ensure that the
lens snout is correctly aligned as described in SOP-3.
3.
Close the sample compartment lid.
In normal use, if the sample compartment cover is opened, then the laser interlock is
breached and laser radiation cannot be emitted from the instrument. Spectrum software
will display the safety interlock is breached Laser Status icon
.
4.
Select the Setup Sample Area XYZ Stage tab, and display the
Live Video
view.
Live Video shows the live image from the video camera.
Содержание RamanStation 400 Series
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanStation 400 Series Getting Started Guide ...
Страница 35: ...Preparing Your Facility ...
Страница 39: ...Installing Your RamanStation 400 Series ...
Страница 52: ...52 RamanStation 400 Series Getting Started Guide ...
Страница 53: ...Introduction to the RamanStation 400 Series ...
Страница 59: ...Fiber Optic Probe ...
Страница 66: ...66 RamanStation 400 Series Getting Started Guide ...
Страница 67: ...Using the RamanStation 400 ...
Страница 91: ...Maintenance and Troubleshooting ...