![Perkin-Elmer RamanStation 400 Series Скачать руководство пользователя страница 69](http://html1.mh-extra.com/html/perkin-elmer/ramanstation-400-series/ramanstation-400-series_getting-started-manual_1530158069.webp)
Using the RamanStation 400 . 69
General SOPs
SOP-1: Turning the Instrument On
1.
Switch on the PC and monitor.
2.
Login to Windows.
3.
Ensure the laser power key is in the ARM position on the rear panel of the instrument.
4.
Ensure the main power supply is connected to a mains outlet and that the outlet is
switched on.
5.
To turn on the spectrometer, press the PSU power switch ON (Figure 32):
- ON
Ο
- OFF
Figure 32 The PSU power switch
6.
Check that the Power LED on the instrument keypad is On.
The Laser LED and the Active LED should be off. If the Power LED is not on, please turn
the instrument off at the PSU and check all of the connections, any power socket
switches and then switch the PSU on again.
7.
If appropriate, connect the stage controller box to a mains outlet and ensure that the
outlet and the controller are switched on.
Содержание RamanStation 400 Series
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanStation 400 Series Getting Started Guide ...
Страница 35: ...Preparing Your Facility ...
Страница 39: ...Installing Your RamanStation 400 Series ...
Страница 52: ...52 RamanStation 400 Series Getting Started Guide ...
Страница 53: ...Introduction to the RamanStation 400 Series ...
Страница 59: ...Fiber Optic Probe ...
Страница 66: ...66 RamanStation 400 Series Getting Started Guide ...
Страница 67: ...Using the RamanStation 400 ...
Страница 91: ...Maintenance and Troubleshooting ...