50 . RamanStation 400 Series Getting Started Guide
Moving the Instrument
WARNING
The RamanStation 400 weighs about 45 kg. Two people are needed to
lift the instrument.
If you need to ship your instrument you must use the original supplied packing materials to
prevent damage to the system and follow the procedures described in this section.
NOTE: When the instrument has reached its destination and has been reassembled, you
should perform the appropriate calibration/verification procedures as described in the
Calibration and Performance Verification SOPs
Motorized XYZ Stage
Before moving your instrument you must fix the stage in position to prevent damage to the
stage.
To prepare the stage:
1.
Remove your sample holder from the stage.
2.
Using the Joystick, or using the Stage Position controls on the Setup Instrument XYZ
Stage tab, move the stage to its lowest position in the Z axis and then roughly to the
center of possible travel in the X and Y axes until the hole for the packing bolt is clearly
visible (see Figure 16 on page 43).
The front and right sides of the stage will be aligned.
3.
Place the packing bolt in the hole in the center of the stage.
4.
Close Spectrum software.
5.
Remove the alignment camera assembly from the sample compartment by removing
the USB cable from the socket labeled
and then unscrewing the lens snout from
the stage.
6.
Tighten the packing bolt using a 5 mm hexagonal key.
Triggered Probe Holster
NOTE: This section is only applicable if you have a RamanStation 400F and have the
optional triggered fiber optic probe holster (L1320297) fitted to your instrument.
Before packing your instrument for shipping, you must remove the triggered fiber optic probe
holster.
Содержание RamanStation 400 Series
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanStation 400 Series Getting Started Guide ...
Страница 35: ...Preparing Your Facility ...
Страница 39: ...Installing Your RamanStation 400 Series ...
Страница 52: ...52 RamanStation 400 Series Getting Started Guide ...
Страница 53: ...Introduction to the RamanStation 400 Series ...
Страница 59: ...Fiber Optic Probe ...
Страница 66: ...66 RamanStation 400 Series Getting Started Guide ...
Страница 67: ...Using the RamanStation 400 ...
Страница 91: ...Maintenance and Troubleshooting ...