Installing Your RamanStation 400 Series . 45
Attaching the External Interlock Override
The RamanStation 400F has two interlock switches, an internal interlock located on the inside
of the sample compartment door, and an external interlock located on the rear panel. To
operate the instrument both interlocks must be satisfied; that is, the sample compartment
door must be closed to collect data, even if data is being collected using the fiber optic
probe.
You must attach the External Interlock Override (L1320264) into the 7 way DIN connector
labeled EXT INTERLOCK on the rear panel of the instrument.
If your laser safety officer determines that an interlock is required on the door to the room,
you will need to purchase the Raman Laser Safety Kit (L1320262). For more information
contact your PerkinElmer representative.
NOTE: If you have purchased a RamanStation 400, then your instrument is a Class 1 laser
product and you do not need to attach an external interlock override.
Содержание RamanStation 400 Series
Страница 1: ...RamanStation 400 Series Getting Started Guide MOLECULAR SPECTROSCOPY ...
Страница 5: ...Introduction ...
Страница 10: ...10 RamanStation 400 Series Getting Started Guide ...
Страница 11: ...Safety Practices ...
Страница 27: ...Raman History Theory and Instrumentation ...
Страница 34: ...34 RamanStation 400 Series Getting Started Guide ...
Страница 35: ...Preparing Your Facility ...
Страница 39: ...Installing Your RamanStation 400 Series ...
Страница 52: ...52 RamanStation 400 Series Getting Started Guide ...
Страница 53: ...Introduction to the RamanStation 400 Series ...
Страница 59: ...Fiber Optic Probe ...
Страница 66: ...66 RamanStation 400 Series Getting Started Guide ...
Страница 67: ...Using the RamanStation 400 ...
Страница 91: ...Maintenance and Troubleshooting ...