Collecting Spectra with the Microscope
.
85
Positioning the Sample
When you use the microscope to collect spectra, you must center the sample on
the sample stage before you collect the background spectrum. This enables you to
set the correct aperture for the sample and then use that aperture when collecting
the background spectrum.
1.
Take out any aperture that is in the retaining well.
2.
Place the sample on the sample stage.
If necessary, secure it with the slide clip:
Figure 18 Parts of the Sample Stage
3.
To set the microscope for viewing a sample, turn the View/IR knob to View.
The detector mirror moves out of the optical path, and the illuminator is
switched on.
4.
Set the Trans/Refl knob to the required position for viewing.
5.
Turn the Illuminator knob to adjust the light to an adequate level for viewing.
NOTE: If you need to increase the illumination during the procedures that follow,
move the knob to a higher setting.
Stage Controls
Slide Clip
Содержание MultiScope System
Страница 1: ...MultiScope System Microscope User s Reference ...
Страница 5: ...Introduction ...
Страница 14: ...14 MultiScope System Microscope User s Reference ...
Страница 15: ...Warnings and Safety Information ...
Страница 26: ...26 MultiScope System Microscope User s Reference Warning Label Near the Lamp Housing Caution hot surface ...
Страница 31: ...Overview of the MultiScope System Microscope ...
Страница 44: ...44 MultiScope System Microscope User s Reference ...
Страница 45: ...Getting Ready to Use the Microscope ...
Страница 53: ...Tutorial Using the Microscope ...
Страница 62: ...62 MultiScope System Microscope User s Reference Figure 12 Reflectance Spectrum of the Coating on a Soda Can ...
Страница 63: ...Preparing Samples ...
Страница 83: ...Collecting Spectra with the Microscope ...
Страница 95: ...Operating the Optional Equipment ...
Страница 109: ...Applications ...
Страница 122: ...122 MultiScope System Microscope User s Reference ...
Страница 123: ...Maintenance ...