value and the instrument will immediately start the experiment keeping the
other programmed parameters in memory.
ENTER 1
∫
1
END METHOD:
YES
•
To ensure that the des red parameter is kept constant, we suggest
to pre-set all other parameters to maximum.
i
s
•
A blinking unit of the controlling parameter appears when the regulation
is complete.
RUN 1
∫
1 99:59h
400V 300mA 50W
•
The timer will alway count down
Voltage
Time
MANUAL PROGRAMMING
Manual programming permits to set voltage, current,
power and time limits for a simple routine electrophoresis run.
Parameters can be changed temporarily without
interrupting the run.
TIMER ACTIVE!
STOP NOW?
10s
1. Verify that the instrument is switched off.
---- STAND-BY ----
MANUAL
2. Connect the electrophoresis tank to the instrument.
3. Switch the instrument on.
If the timer is still running after switching on, the power supply will
automatically proceed the experiment for the remaining time.
Press STOP (with n 10 seconds) to enter the stand by mode to
eventually change parameters. Press RUN to proceed with the experiment
or pres SET for 2 seconds to re-enter the stand by mode.
i
-
s
-
ENTER
VOLTAGE: 400V
4. Select the MANUAL mode and press SET.
ENTER
CURRENT: 300mA
5. Select the desired voltage limit and press SET.
6. Select the desired current limit and press SET.
7. Select the desired power limit and press SET.
8. Enter the desired time-units (hours or kilovolt hours) and
press SET
ENTER
POWER: 50W