2
0
2
2
-0
2
40
Configuration
General
0xCC0001
Custom Name Char[32] r/w
User-defined name
0x510001
AutoTrigger
0.1
r/w
Deactivation/activation of
cyclical image capture.
0x680001
Exposure time 1 .. 1000
0
r/w
Sets the manual exposure
time. To set the exposure
time manually, the "Use
manual exposure time"
(see "Illumination Menu
Item" on page 54) function
must be activated. Increas-
ing the value increases the
exposure time and there-
fore the image brightness.
Values below 1000
s are
suitable in most cases.
0xBE0001
Use manual
exposure time
0.1
r/w
When enabled, the manu-
ally set exposure time is
used. If this box is not
checked, the exposure
time is regulated automati-
cally during the "teach-in"
process
0x100001
Flash time
1 .. 1000
0
r/w
Exposure time of used
flash light when triggering
LEDs in
µ
s (see
0x640002)
0x9F0001
Object con-
trast
1 .. 100
r/w
Contrast threshold for
detecting laser line on
object in % (default value:
45)
0xC90001
ROI min X
-
3200 ...
3200
r/w
The lowest value on the x-
axis of the ROI (in mm).
0xC90002
ROI max X
-
3200 ...
3200
r/w
The highest value on the x-
axis of the ROI (in mm).
0xC90003
ROI min Z
1 ...
3200
r/w
The lowest value on the z-
axis of the ROI (in mm).
0xC90004
ROI max Z
1 ...
3200
r/w
The highest value on the z-
axis of the ROI (in mm).
ImageTrans-
ferActive
Transfer
images
0.1
r/w
Automatically transfer
images.
NOTE: may increase anal-
ysis time.
Config ID
Parameter ID
Parameter
name
Values
read (r)/
write (w)
Description