Pulscon LTC57 PROFIBUS PA
Operating menu
15.3.12
Switch output submenu
Switch output
Switch output function
Navigation
Setup
Advanced setup
Switch output
Blocking
Navigation
Setup
Advanced setup
Switch output
Switch out funct
Description
Select function for switch output.
Selection
• Off
• On
• Diagnostic behavior
• Limit
• Digital output
Factory setting
Off
Additional information
Meaning of the options
•
Off
The output is always open (non-conductive).
•
On
The output is always closed (conductive).
•
Diagnostic behavior
The output is normally closed and is only opened if a diagnostic event is
present. The
Assign diagnostic behavior
parameter (page 133) determines
for which type of event the output is opened.
•
Limit
The output is normally closed and is only opened if a measured variable
exceeds or falls below a defined limit. The limit values are defined by the
following parameters:
–
Assign limit
–
Switch-on value
–
Switch-off value
•
Digital output
The switching state of the output tracks the output value of a DI function block.
The function block is selected in the
Assign status
parameter (page 133).
Note:
The
Off
and
On
options can be used to simulate the switch output.