201
5-0
3
122
Pulscon LTC57 PROFIBUS PA
Operating menu
Table number
Level (Manual)
Level (Semi-automatic)
Customer value
Blocking
Navigation
Setup
Advanced setup
Linearization
Table number
Prerequisite
Linearization type
Table
Description
Select table point you are going to enter or change.
User entry
1 to 32
Factory setting
1
Blocking
Navigation
Setup
Advanced setup
Linearization
Level
Prerequisite
•
Linearization type
(page 116) =
Table
•
Table mode
(page 121) =
Manual
Description
Enter level value of the table point (value before linearization).
User entry
Signed floating-point number
Factory setting
0 %
Navigation
Setup
Advanced setup
Linearization
Level
Prerequisite
•
Linearization type
(page 116) =
Table
•
Table mode
(page 121) =
Semi-automatic
Description
Displays measured level (value before linearization). This value is transmitted to the
table.
Blocking
Navigation
Setup
Advanced setup
Linearization
Customer value
Prerequisite
Linearization type
Table
Description
Enter linearized value for the table point.
User entry
Signed floating-point number
Factory setting
0 %