2597
78 2013-
02
110
Commands
10.5
Special Command Modes for IPC03 Read/Write Tags
IPC03 Configuration
The storage of a data carrier IPC03 is organized by word. A data word is defined
with a length of 32 bits. For the normal data range, 29 words from addresses 3
through 31 (<WordAddr> = 00h ... 1Ch) are available.
Word 0 contains the password. The password can only be written.
With word 1, the "Protection Word", you can define a read-protected and a write-
protected range. The "Protection Word" can only be read and written with the
correct password.
With word 2, the "Control Word", you can set various operating modes and the
read range for the operating mode "Default Read". The "Control Word" can only
be read and written with the correct password.
If you would like to use the "Protection Word" and the "Control Word", you must
first activate the password mode.
4
falscher Befehl
command error
5
kein Transponder
no tag
6
Hardwarefehler
hardware error
7
Softwarefehler
software error
8
Service-Schnittstelle
service interface
9
unbekannter Fehler
unknown error
Status
German display text
English display text
Note!
You can only use the commands in this section for the data carrier type '03'
(IPC03).
Address
Meaning
<WordAddr> <ConfAddr> Note
Word 0
Password
-
-
Write only
Word 1
Protection word
-
1
Read/write
Word 2
Control word
-
2
Read/write
Word 3 ...31
Data range
00h ... 1Ch
-
Read/write
Word 32
Device Serial Number
1Dh
-
Read only
Word 33
Device identification
1Eh
-
Read only