February 2011 

 

 

SWITCHING ON 

 

 

 

Before installing the CWT and taking a measurement refer to SAFETY AND PRE-USE 
CHECKS to ensure safe operation of your CWT  

 

 

Do not fit a Rogowski coil to a live circuit; always de-energise the circuit first. 

 
1.  Connect  the  BNC  plug  on  the  output  cable  of  the  transducer  to  your  oscilloscope  or  current 

monitoring equipment. The CWT must  only be used with oscilloscopes or monitoring equipment 
which have their BNC INPUTS PROPERLY GROUNDED. 

 
2.  Having  carried  out  the  VISUAL  INSPECTION  of  the  Rogowski  coil,  un-clip  the  coil  and  wrap  it 

around the de-energised conductor under test. 

 
3.  Insert the free-end of the coil fully inside the ferrule and lock the jaws if necessary. 
 
4.  Re-energise the conductor.  

 

5.  The CWT is switched ON by pressing and releasing the ON push button, and is turned OFF by 

depressing the button fully; the LED indicates that the CWT is ON when the LED is GREEN. 

 

6. 

– EITHER -

 Standard alkali battery supplied CWT units

 

- B

 

 
Four  1.5V  AA  alkali  batteries  provide  about  70  hours  operation.  Battery  voltage  is  continuously 
monitored; healthy batteries are indicated by the GREEN LED. If the LED is RED the batteries are 
delivering  less  than  2V  and  must  be  replaced  by  removing  the  sliding  battery  door  in  the  back 
cover of the integrator enclosure. 
 

 

The units can also be powered by an external DC supply. The DC voltage can be between  
12V(-10%) and 24V (+10%). When the DC supply is present a RED indicating LED adjacent to 
the socket is illuminated. With the DC supply present the batteries are inoperative. 
 

- OR -

  Re-chargeable battery CWT units - 

R

 

 

When fully charged four 1.2V AA NiMH rechargeable cells provide about 30hrs operation. Battery 
voltage is continuously monitored; healthy batteries are indicated by the GREEN LED. If the LED 
is RED the batteries are delivering less than 2V and must be recharged. 
 
The units can also be powered by an external DC supply. The DC voltage can be between  
12V(-10%) and 24V (+10%). When the DC supply is present a RED indicating LED adjacent to 
the socket is illuminated. When the DC supply is present the batteries are inoperative and the 
external DC voltage powers the integrator. In addition when the DC supply is present (regardless 
of whether the integrator is on or off) the rechargeable batteries are trickle charged.  

 
7.  After switch-on the CWT requires a settling down time to attain its quiescent state before 

providing correct current measurement. The time, which depends on warm-up and low frequency 
bandwidth, can be as long as 2 minutes. 

Содержание CWT

Страница 1: ...ELECTRONIC MEASUREMENTS Ltd Tel 44 115 946 9657 Fax 44 115 946 8515 Email info pemuk com Web www pemuk com February 2011 Distributed By 955 Industrial Road San Carlos CA 94070 USA GMW Associates PHONE 1 650 802 8292 FAX 1 650 802 8298 EMAIL sales gmw com WEB www gmw com ...

Страница 2: ... present the batteries are inoperative Battery type 4 four 1 5V AA alkali Battery life 70 hrs typ DC socket type 2 1 or 2 5mm jack socket polarity indicated on front of CWT DC voltage 12V 10 to 24V 10 DC quiescent supply current 30mA 12V DC 30mA 24V DC When specified with rechargeable batteries the CWT operates directly from the batteries when there is no DC supply present and is powered directly ...

Страница 3: ...m for closing the Rogowski coil 4 Cable connecting the Rogowski coil to the integrator electronics 5 Socket for external DC supply 6 LED status indicator for external DC supply 7 Enclosure for the integrator electronics 8 Output BNC socket 9 4 x AA batteries 10 LED indicator GREEN CWT is ON RED low battery 11 Push button ON OFF switch ...

Страница 4: ...e at 15kVrms using a 50Hz sinewave Coils are tested without the 10kV sleeve fitted The test also covers the ferrule connecting the coil to the cable For additional safety it is recommended that the ferrule is not situated in direct contact with conductors at voltages higher than 8kV peak See also Obtaining the Best Measurement The voltage ratings are appropriate for intermittent use of the CWT as ...

Страница 5: ... to vibration the locking nut should be engaged as shown in the diagram below To lock the jaws The locking mechanism must not be actuated at temperatures below 0o C Care must be taken not to over tighten the locking nut THE INTEGRATOR The CWT must only be used with oscilloscopes or monitoring equipment which have their BNC INPUTS PROPERLY GROUNDED ...

Страница 6: ...are delivering less than 2V and must be replaced by removing the sliding battery door in the back cover of the integrator enclosure The units can also be powered by an external DC supply The DC voltage can be between 12V 10 and 24V 10 When the DC supply is present a RED indicating LED adjacent to the socket is illuminated With the DC supply present the batteries are inoperative OR Re chargeable ba...

Страница 7: ...ncy oscillations in the MHz range then measurement error can arise due to capacitive coupling to the coil As a check on any unwanted response to adjacent fields it is wise to display the output of the CWT when close to but not encircling the conductor whose current is to be measured This will reveal the magnitude of any unwanted response to currents close to but outside the coil EMC EMISSIONS PEM ...

Страница 8: ...of repair or replacement RETURNING CWTs FOR RECALIBRATION OR REPAIR If it is necessary to return the CWT to PEM for repair whether or not under warranty please contact PEM in advance for shipping instructions see below for contact details To facilitate customs clearance it is important to follow the correct procedure otherwise import tax will be charged PEM will not accept units sent for recalibra...

Отзывы: