INTRODUCTION 

 
PEM‟s Rogowski Current Waveform Transducers combine high bandwidth, safety, and the minimum 
disruption to the circuit under test. These instructions should be followed whenever the unit is used. 
They are intended to help you obtain the best and safest performance from the transducer. 

 
 

SPECIFICATION 

 

ACCURACY (5 to 100% full scale) 

 
Within the limits of bandwidth and low frequency noise that are specified on the relevant CWT 
specification sheet. 
 
Calibration: 

Normally ±0.2% (traceable to UKAS) with the 
conductor central in the CWT loop. See the calibration 
certificate for more details. 

Positional accuracy: 

±1% typ. variation of accuracy with conductor position 
in the loop (see „Obtaining The Best Measurement‟) 

Linearity: 

±0.05% typ. 

 

POWER SUPPLY 

 
The CWT can be powered by both battery 

and

 an external DC voltage. With the external DC supply 

present the batteries are inoperative. 
 
Battery type: 

4 (four) 1.5V AA alkali  

Battery life: 

70 hrs typ. 

DC socket type: 

2.1 or 2.5mm jack socket 

– polarity indicated on front of 

CWT 

DC voltage: 

12V(-10%)  to 24V(+10%) 

DC quiescent supply current: 

30mA (@12V DC)  ,  30mA (@ 24V DC) 

 
When specified with 

rechargeable batteries

 the CWT operates directly from the batteries when there 

is no DC supply present and is powered directly from the DC supply when present. With the DC 
supply present the rechargeable batteries are permanently trickle charged. 
 
Battery type: 

4 (four) 1.2V NiMH rechargeable batteries 

Battery life: 

30 hrs typ. (based on a 1400mAH cell) 

Recharge time: 

40 hrs typ. 

DC socket type: 

2.1 or 2.5mm jack socket 

– polarity indicated on front of 

CWT 

DC voltage: 

12V(-10%)  to 24V(+10%) 

DC quiescent supply current: 

70mA (@12V DC)  ,  70mA (@ 24V DC) 

 

OUTPUT 

 
Maximum output voltage: 

±6V (corresponding to ±Peak Current Rating of CWT) 

Output cable: 

0.5m BNC to BNC 50Ω cable 

Minimum output loading: 

100kΩ 

 

OPERATING TEMPERATURE RANGES 

 
Rogowski coil and cable 

-20

o

C to 100

o

Integrator electronics 

0 to 40

o

Содержание CWT

Страница 1: ...ELECTRONIC MEASUREMENTS Ltd Tel 44 115 946 9657 Fax 44 115 946 8515 Email info pemuk com Web www pemuk com February 2011 Distributed By 955 Industrial Road San Carlos CA 94070 USA GMW Associates PHONE 1 650 802 8292 FAX 1 650 802 8298 EMAIL sales gmw com WEB www gmw com ...

Страница 2: ... present the batteries are inoperative Battery type 4 four 1 5V AA alkali Battery life 70 hrs typ DC socket type 2 1 or 2 5mm jack socket polarity indicated on front of CWT DC voltage 12V 10 to 24V 10 DC quiescent supply current 30mA 12V DC 30mA 24V DC When specified with rechargeable batteries the CWT operates directly from the batteries when there is no DC supply present and is powered directly ...

Страница 3: ...m for closing the Rogowski coil 4 Cable connecting the Rogowski coil to the integrator electronics 5 Socket for external DC supply 6 LED status indicator for external DC supply 7 Enclosure for the integrator electronics 8 Output BNC socket 9 4 x AA batteries 10 LED indicator GREEN CWT is ON RED low battery 11 Push button ON OFF switch ...

Страница 4: ...e at 15kVrms using a 50Hz sinewave Coils are tested without the 10kV sleeve fitted The test also covers the ferrule connecting the coil to the cable For additional safety it is recommended that the ferrule is not situated in direct contact with conductors at voltages higher than 8kV peak See also Obtaining the Best Measurement The voltage ratings are appropriate for intermittent use of the CWT as ...

Страница 5: ... to vibration the locking nut should be engaged as shown in the diagram below To lock the jaws The locking mechanism must not be actuated at temperatures below 0o C Care must be taken not to over tighten the locking nut THE INTEGRATOR The CWT must only be used with oscilloscopes or monitoring equipment which have their BNC INPUTS PROPERLY GROUNDED ...

Страница 6: ...are delivering less than 2V and must be replaced by removing the sliding battery door in the back cover of the integrator enclosure The units can also be powered by an external DC supply The DC voltage can be between 12V 10 and 24V 10 When the DC supply is present a RED indicating LED adjacent to the socket is illuminated With the DC supply present the batteries are inoperative OR Re chargeable ba...

Страница 7: ...ncy oscillations in the MHz range then measurement error can arise due to capacitive coupling to the coil As a check on any unwanted response to adjacent fields it is wise to display the output of the CWT when close to but not encircling the conductor whose current is to be measured This will reveal the magnitude of any unwanted response to currents close to but outside the coil EMC EMISSIONS PEM ...

Страница 8: ...of repair or replacement RETURNING CWTs FOR RECALIBRATION OR REPAIR If it is necessary to return the CWT to PEM for repair whether or not under warranty please contact PEM in advance for shipping instructions see below for contact details To facilitate customs clearance it is important to follow the correct procedure otherwise import tax will be charged PEM will not accept units sent for recalibra...

Отзывы: