8
047657 r7
MaIn ConTrol boarD
DIagnosTIC DIsPlaY
0
1
2
3
4
No input received
Foot switch button pressed
Touch pad button pressed
Foot switch and touch pad button pressed
Chair stopped due to loss of input signal
0
1
2
3
4
5
6
7
8
No motion in progress
Backrest in motion
Base in motion
Backrest and loose in motion
Chair stopped due to no sensed motion
Backrest input received, out of soft limits
Base input received, out of soft limits
Backrest and base input received,
out of soft limits
Auto position in progress
No output present
Normal operation
Normal operation
Normal operation
Examine potentiometer
Reprogram soft limits
Reprogram soft limits
Reprogram soft limits
Normal operation
00
FF
E_
RICHT DECIMAL
LEFT DECIMAL
DenTal CHaIr DIagnosTIC CoDes
InPuT
sTaTus
DIgIT
InPuT sTaTus
CorreCTIVe aCTIon
Examine input device
Switch may be stuck
Switch maybe stuck
Multiples switches stuck
Input device or connection faulty
OUTPUT STATUS DIGIT
OUTPUT STATUS
CORRECTIVE ACTION
DIAGNOSTIC DISPLAY
OPERATING STATUS
CORRECTIVE ACTION
Potentiometer value too low
Potentiometer value too high
Main control board failure (replace)
Error is for base potentiometer
Error is for back rest potentiometer
Calibrate potentiometer
Calibrate potentiometer
Replace board
Examine potentiometer
Examine potentiometer
Содержание Spirit 3000 Series
Страница 1: ...Pelton and Crane Series 3000 Dental Chair Collins Bay Institution Warehouse ...
Страница 2: ...SPIRIT 3000 CHAIR TECHNICAL DIMENSIONAL DATA ...
Страница 3: ...SPIRIT 3000 CHAIR Dimensions w Junction Box Plumbing Diagram ...
Страница 5: ...Spirit 3000 Series Dental Chair Installation Instructions Model 3001 Model 3004 Model 3003 Model 3002 ...
Страница 11: ...7 047657 r7 ...
Страница 16: ...Spirit 3000 Series Dental Chair USE AND CARE Model 3001 Model 3004 Model 3002 Model 3003 ...