background image

 

EMC GUIDANCE

 

150 kHz to 80 MHz 

80 MHz to 800 MHz 

800 MHz to 2.5 GHz 

Table 6 Recommended separation distances between portable and mobile 

RF communications equipment and the ME EQUIPMENT or ME SYSTEM –
for ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING 

Recommended separation distances 

between  portable and mobile RF communications equipment and the device.

The device is intended for use in an electromagnetic environment in which 

radiated RF disturbances are controlled. The customer or the user of the device

can help prevent electromagnetic interference by maintaining a minimum distance 

between portable and mobile RF communications equipment (transmittters) and the 

device as recommended below, according to the maximum output power of the 

communications equipment.

Rated maximum output

power of transmitter

(W) 

Separation distance according to frequency of transmitter (m)

For transmitters rated at a maximum output power not listed above, the recommended 

separation distance 

d

 in metres (m) can be estimated using the equation applicable to 

the frequency of the transmitter, where 

P

 is the maximum output power rating of the 

transmitter in watts (W) according to the transmitter manufacturer.

NOTE 1    At 80MHz and 800MHz, the separation distance for the higher frequency 

                 range applies.

NOTE 2    These guidelines may not apply in all situations. Electromagnetic propagation

                 is affected by absorption and reflection from structures, objects and people.

0.01

0.1

1

10

100

0.12

0.12
0.38
1.2
3.8
12

0.23
0.73
2.3
7.3
23

=

d

P

=

d

1,2

2,3

P

=

d

0.38
1.2
3.8
12

P

1,2

25

Содержание Vena TMB-988

Страница 1: ...oosing the Vena Blood Pressure Monitor To use the monitor correctly and safely please read the manual thoroughly Please keep this manual well in order to reference in future AVG SYS kPammHg AVG DIA PM...

Страница 2: ...BLOOD PRESSURE What are systolic pressure and diastolic pressure What is the standard blood pressure classification Why does my blood pressure fluctuate throughout the day Why do I get a different bl...

Страница 3: ...ssure of children Ask your doctor before using it on older children The device is not intended for patient transport outside a healthcare facility The device is not intended for public use This device...

Страница 4: ...essure reaches 40 kPa 300 mmHg the unit will automatically deflate Should the cuff not deflate when pressures reaches 40 kPa 300 mmHg detach the cuff from the wrist and press the START STOP button to...

Страница 5: ...an inaccurate measurement Blood pressure monitor is detecting an irregular heartbeat during measurement Irregular heartbeat Blood pressure level indicator Indicate the blood pressure level Heartbeat...

Страница 6: ...your local recycling guidelines Install the batteries as indicated in the battery compartment Always use the correct battery type 2 AAA batteries Setting Date Time and Measurement Unit 2 Press MEM bu...

Страница 7: ...rtably with legs uncrossed feet flat on the floor back and wrist supported Rest for 5 minutes before measuring Wait at least 3 minutes between measurements This allows your blood circulation to recove...

Страница 8: ...ent record Press MEM button or SET button to get the record you want The most recent record 1 is shown first Each new measurement is assigned to the first 1 record All other records are pushed back on...

Страница 9: ...ut deleting any records press START STOP button before pressing SET button to confirm any delete commands 2 Press SET button to confirm deleting the LCD will display dEL dOnE and then turn off Tips fo...

Страница 10: ...en we recommend you seek medical advice Please note that the device does not replace a cardiac examination but serves to detect pulse irregularities at an early stage The chart on the right is the sta...

Страница 11: ...rther assistance Refer to the warranty for contact information and return instructions The measurement process does not detect the pulse signal Loosen the clothing on the wrist and then measure again...

Страница 12: ...on Type BF applied part Protection against ingress of water Software version Device classification V01 Internally Powered ME Equipment WARNING No modification of this equipment is allowed Temperature...

Страница 13: ...evaluation of medical devices Part 10 Tests for irritation and skin sensitization EN ISO 15223 1 2016 ISO 15223 1 2016 Medical devices Symbols to be used with medical device labels labeling and inform...

Страница 14: ...claration electromagnetic immunity IMMUNITY test Not applicable 3 V m Compliance level IEC 60601 TEST LEVEL Conducted RF IEC 61000 4 6 Radiated RF IEC 61000 4 3 3 V m 80 MHz to 2 5 GHz 3 Vrms 150 kHz...

Страница 15: ...device as recommended below according to the maximum output power of the communications equipment Rated maximum output power of transmitter W Separation distance according to frequency of transmitter...

Отзывы: