TOC 1000

User Manual

Copyright © Peak Scientific Instruments Ltd 2021 — TOC 1000 User Manual  Rev 2  RSID 3471 EN —31/08/2021

Содержание TOC 1000

Страница 1: ...TOC 1000 User Manual Copyright Peak Scientific Instruments Ltd 2021 TOC 1000 User Manual Rev 2 RSID 3471 EN 31 08 2021...

Страница 2: ...ing 10 Fittings Kit Contents 11 Installation 12 Generator Environment 12 Generator Overview 13 General Dimensions 13 Rear Connections 14 Electrical Connection 15 Air Purity 16 Class 1 Particulate 16 C...

Страница 3: ...nce document where you can skip to the relevant information Users can refer to the contents page to find the relevant information Please review each of the following sections carefully Thank you for s...

Страница 4: ...upply to the generator must comply with ISO 8573 1 2010 Class 1 2 1 at all times d your air compressor dryer filtration and oil removal systems must be deemed suitable for use by Peak or a Peak Partne...

Страница 5: ...nd conditions outlined in this warranty state ment in any respect or to create any additional obligations or liabilities for any party involved 6 This warranty statement supersedes any and all prior w...

Страница 6: ...ted conditions are fully understood or met A CAUTION notice denotes a hazard It calls attention to an operating procedure process or similar which if not correctly performed or adhered to could cause...

Страница 7: ...61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use Electromagnetic Compatibility Directive 2014 30 EU EN 61326 1 2013 Electrical Equipment for Measure...

Страница 8: ...ent Safety Regulations 2016 SI 2016 1101 as amended BS61010 1 2010 Safety Requirements for Electrical Equipment for Measurement Control and Laboratory Use The Electromagnetic Compatibility Regulations...

Страница 9: ...Guidance Notes PDF produced by the Department for Business Innovation and skills for the UK and here for Europe All PEAK products that are subject to the WEEE directive are compliant with the WEEE mar...

Страница 10: ...hthalates Phthalate BHT Free Suspended Liquids None Gas Outlets 1 x 1 4 BSPP x 6mm push fit connection Start Up Time 120 minutes Electrical Requirements Voltage 120V 5 230V 10 Frequency 120V 60Hz 230V...

Страница 11: ...e Peak Partner from where the unit was purchased Follow the unpacking instructions posted on the side of the crate It will require two people to remove the unit from the shipping crate and to manoeuvr...

Страница 12: ...ings Kit are as follows 1 Tube PE 6mmOD 4mmID x3m 2 Cable C13 BS 1363 UK 230V 10A 2m x1 3 Cable C13 Nema 6 15 US 220V 2m x1 4 Cable C13 Schuko H05VV F EU 230v 2 5m x1 5 Cable C13 Nema 5 15 US 110V 2m...

Страница 13: ...ing systems can sometimes give rise to pockets of air with high relative humidity Operation of the unit within such a pocket could adversely affect its performance Consideration should also be given t...

Страница 14: ...Page 14 Generator Overview General Dimensions The generator must always be placed on a flat level surface Failure to do so will affect the performance of the generator WARNING 270 mm 343 mm 530 mm...

Страница 15: ...connected to correct sources and applications All Connections should only be carried out by trained personnel Generator must be switched off and unplugged prior to any cleaning or maintenance operatio...

Страница 16: ...ratings and appropriate approvals for the country of operation Failure to do so could cause damage to the generator or risk of overloading of the power cord This unit is classified as SAFETY CLASS 1 T...

Страница 17: ...late count should not exceed 20 000 particles in the 0 1 0 5 micron size range 400 particles in the 0 5 1 micron range and 10 particles in the 1 5 micron size range Class 4 Water A pressure dew point...

Страница 18: ...first connected to the application the generator should be operated in isolation i e not connected to the application for 120 minutes This is to ensure that any impurities are purged from the system F...

Страница 19: ...and gas purity Tubing Lengths The diameter of the tubing which will be connected to the gas outlet is important and is determined by the length of tubing required Failure to follow these recommendati...

Страница 20: ...20 Service Requirements Service Schedule Purchase Interval Component Qty Visit 12 Months TOC 1000 Annual Maintenance Kit 1 www peakscientific com ordering Note Only Peak approved components should be...

Страница 21: ...to install and setup your generator IQ OQ Certified assurance for applications requiring documented qualification Complete plan Swift response by a Peak Service Engineer within 72 hours planned preve...

Страница 22: ...he cloth prior to use Cleaning should only be undertaken with the power switched off and the power cord removed from the rear of the generator Under no circumstances should any solvents or abrasive cl...

Страница 23: ...Page 23 Troubleshooting Problem Possible Solution The application is reporting low pressure Check for leaks between the TOC 1000 and the application Contact your service provider...

Страница 24: ...ic com For generators that remain unregistered after 1 month from the shipment date the warranty will be considered active from the date of factory dispatch Complete Plan only Go Online or Complete an...

Страница 25: ...Page 25...

Страница 26: ...iority To discuss Peak Protected generator cover and payment options speak to your local Peak Representative or for further information contact protected peakscientific com For further information on...