Page 8
8.
7.
Unlock the generator right side panel, use the
8mm Hex Key 00-0007
,
and turn the
lock 90° to the right.
00-0007
Connect the
6mm PE Tubing 00-1282
to the
¼”- 6mm Push Fit Fitting 02-4600
.
Fit the other end of the tubing to a suitable drain connection or container. The
container
MUST NOT
have an airtight seal as water and air will be expelled
periodically under pressure.