Troubleshooting 

 

Problem 

Possible solution 

The viscometer does not work 

Check the batteries 

The instrument does not show “zero” without any 
fluids  

Use the auto zero function 

Viscosity speed is unstable and inaccurate 

Check the rheological characteristics of the fluid 
Perform a calibration check by using calibration 
standards 
Check if the rotor disk rotates steadily 
Check if the temperature of the sample is stable 

 

Further information 

9.1 

Calibration 

The viscometer comes factory-calibrated and with a calibration certificate.  
 
We recommend checking the accuracy of the viscometer on a regular basis by using optional calibration 
oils. 
 
For the verification, we recommend using the R3 rotor and a calibration standard of 1,000 mPa s. Perform 
at least 5 measurements and compare the reading to the viscosity of the calibration oil. 
 

Note: Keep in mind that the viscosity is greatly influenced by the temperature. Therefore, you should 
perform the verification under the specified conditions of the calibration oil. 

9.2 

Reference substances and viscosities 

 

Substance 

Approx. viscosity (mPas) 

Motor oil SAE 10 

65 

Olive oil 

84 

Coatings (airbrushed) 

100 

Yoghurt 

150 

Sugar solution 70 % 

400 

Lubricating oil 

50 

– 1,000 

Concentrated juice 

1,500 

Inks 

550 

– 2,200 

Honey 

10,000 

Nanocellulose (common additive) 

8,000 

– 10,000 

Toothpaste 

100,000 

 
 

 

Содержание PCE-RVI 7

Страница 1: ...truments UK Ltd Units 12 13 Southpoint Business Park Ensign way Hampshire Southampton United Kingdom SO31 4RF From outside UK 44 Tel 0 2380 98703 0 Fax 0 2380 98703 9 info pce instruments com www pce instruments com english www pce instruments com Manual Viscometer PCE RVI 7 Version 1 0 Date of creation 16 11 2015 Date of last change 07 06 2016 ...

Страница 2: ...nsert replace the batteries 6 5 2 Attach a rotor spindle 6 5 3 Measuring vessel 6 6 Measurement 7 6 1 Turn on the viscometer 7 6 2 Select a rotor spindle 7 6 3 Take a measurement 7 7 Further functions 8 7 1 Auto zero 8 7 2 Change the measuring unit 8 8 Troubleshooting 9 9 Further information 9 9 1 Calibration 9 9 2 Reference substances and viscosities 9 10 Disposal 10 11 Contact 10 11 1 PCE Instru...

Страница 3: ...nel of PCE Instruments The instrument should never be placed with the user interface facing an object e g keyboard side on a table You must not make any technical changes to the device The appliance should only be cleaned with a damp cloth use only pH neutral cleaner Do not expose the device or its components except for the rotors to water or other liquids Do not operate the device with wet hands ...

Страница 4: ...R6 or R6 batteries 6 V Battery life 24 30 hours of continuous operation Operating conditions 10 40 C 80 RH non condensing max 2 000 m above mean sea level Pollution degree 2 Protection class IP 20 Dimensions 170 x 110 x 410 mm Weight 1 8 kg Viscosity table Rotor Measuring range dPa s Measuring range mPa s Measuring vessel Sample capacity R3 0 3 13 dPa s 30 1 300 mPa s No 3 170 ml R1 3 150 dPa s 30...

Страница 5: ...Manual 5 4 System description 4 1 Button and display description Attached rotor Battery status Full scale percentage Down button Enter button Up button Viscosity Rotational speed ...

Страница 6: ...y unit and carefully turn it clockwise Note Be careful when attaching the rotor to prevent damage to the rotor and shaft Note Always keep the engine shaft and rotors clean especially the screw connection Note Every rotor has a mark with its name on it at the top end 5 3 Measuring vessel The included measuring vessel has 3 pins on the bottom side which fit into the 3 holes on the base plate of the ...

Страница 7: ... button you can go to the next or previous rotor Once you have found the attached rotor press Enter to confirm the selection 6 3 Take a measurement First please make sure that the viscometer is levelled properly Next use the height adjustment at the side of the viscometer and adjust it so that the measuring vessel can be placed on the base plate without touching the rotor After you placed the vess...

Страница 8: ... start the zeroing procedure During the zeroing the display shows the following screen After the zeroing was successful the viscometer automatically goes back to the information screen 7 2 Change the measuring unit To select the measuring unit you can choose between mPa s and dPa s turn on the viscometer by pressing and holding the Enter button for 6 to 10 seconds Once the information screen appea...

Страница 9: ...g the accuracy of the viscometer on a regular basis by using optional calibration oils For the verification we recommend using the R3 rotor and a calibration standard of 1 000 mPa s Perform at least 5 measurements and compare the reading to the viscosity of the calibration oil Note Keep in mind that the viscosity is greatly influenced by the temperature Therefore you should perform the verificatio...

Страница 10: ... use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 11 Contact If you have any questions about our range of products or measuring instruments please contact PCE Instruments 11 1 PCE Instruments UK By post PCE Instruments UK Ltd Units 12 13 Southpoint Business Park Ensign Way Southampton Hampshire Unit...

Страница 11: ...4 D 59872 Meschede Deutschland Tel 02903 976 99 0 Fax 02903 976 99 29 info pce instruments com www pce instruments com deutsch Bedienungsanleitung Viskosimeter PCE RVI 7 Version 1 0 Erstelldatum 16 11 2015 Letzte Änderung 14 01 2016 ...

Страница 12: ...stenbeschreibung 5 5 Messvorbereitungen 6 5 1 Batterien einsetzen austauschen 6 5 2 Messspindel montieren 6 5 3 Messbehälter 6 6 Messen 7 6 1 Viskosimeter einschalten 7 6 2 Messspindel auswählen 7 6 3 Messung durchführen 7 7 Weitere Funktionen 8 7 1 Auto Zero 8 7 2 Auswahl der Maßeinheit 8 8 Fehlerbehebung 9 9 Weitere Informationen 9 9 1 Kalibrierung 9 9 2 Substanzen und ihre Viskosität 9 10 Entso...

Страница 13: ...lifiziertes Personal der PCE Deutschland GmbH vorgenommen werden Vermeiden Sie jeglichen Kontakt des Gerätes und seiner Komponenten außer der Messspindeln mit Wasser oder sonstigen Flüssigkeiten Bedienen Sie das Gerät nicht mit nassen Händen Benutzen Sie zum Reinigen des Gerätes ein feuchtes Tuch Benutzen Sie keine aggressiven Reinigungs oder Scheuermittel Legen Sie das Gerät nicht mit dem Display...

Страница 14: ...2 R3 Stromversorgung 4 x AA LR6 oder R6 Batterien 6 V Batterielaufzeit 24 30 Stunden im Dauerbetrieb Betriebsbedingungen 10 4 C 80 r F nicht kondensierend max 2 000 m über dem Meeresspiegel Verschmutzungsgrad 2 Schutzklasse IP 20 Abmessungen 170 x 110 x 410 mm Gewicht 1 8 kg Viskositätstabelle Rotor Messbereich dPas Messbereich mPas Messbecher Kapazität R3 0 3 13 dPas 30 1 300 mPas Nr 3 170 ml R1 ...

Страница 15: ...BETRIEBSANLEITUNG 5 4 Systembeschreibung 4 1 Display und Tastenbeschreibung Ausgewählte Messspindel Batteriestatus des Messbereiches Ab Taste Enter Taste Auf Taste Viskosität Umdrehungs geschwindigkeit ...

Страница 16: ...eien Sie beim Anschrauben der Messspindeln äußerst vorsichtig um Beschädigungen an der Spindel und Welle z B durch Verbiegen zu vermeiden Hinweis Halten Sie die Messspindeln und die Motorwelle stets sauber dies gilt besonders für die Schraubverbindung Hinweis Jede Messspindel hat eine Markierung mit ihrer Bezeichnung am oberen Ende 5 3 Messbehälter Der mitgelieferte Messbehälter verfügt über drei ...

Страница 17: ...igt wird Mit der Auf bzw Ab Taste gelangen Sie zur nächsten bzw vorherigen Spindel Mit der Enter Taste können Sie die Auswahl bestätigen 6 3 Messung durchführen Stellen Sie vor einer Messung sicher dass das Viskosimeter korrekt ausgerichtet ist Benutzen Sie anschließend die Höhenverstellung an der Seite des Gerätes und stellen Sie diese so ein dass der Messbehälter auf der Basisplatte platziert we...

Страница 18: ...m den Vorgang zu starten Während der Auto Zero Vorgang läuft sehen Sie folgenden Bildschirm Nachdem der Vorgang erfolgreich abgeschlossen wurde werden Sie automatisch zum Informationsbildschirm weitergeleitet 7 2 Auswahl der Maßeinheit Um die Maßeinheit zu ändern von dPas zu mPas oder umgekehrt schalten Sie das Messgerät ein indem Sie die Enter Taste für 6 bis 10 Sekunden gedrückt halten Wenn der ...

Страница 19: ...ie Genauigkeit des Viskosimeters in regelmäßigen Abständen zu überprüfen Benutzen Sie dazu unsere optionalen Kalibrieröle Für die Überprüfung der Kalibrierung empfehlen wir die Verwendung der R3 Messspindel in Verbindung mit einem Kalibrieröl mit einer Viskosität von 1 000 mPas Führen Sie mindestens 5 Messungen durch und vergleichen Sie anschließend die Ergebnisse mit den Eigenschaften des Kalibri...

Страница 20: ...tV PCE Deutschland GmbH Im Langel 4 59872 Meschede Zur Umsetzung der ElektroG Rücknahme und Entsorgung von Elektro und Elektronikaltgeräten nehmen wir unsere Geräte zurück Sie werden entweder bei uns wiederverwertet oder über ein Recyclingunternehmen nach gesetzlicher Vorgabe entsorgt 11 Kontakt Bei Fragen zu unserem Produktsortiment oder dem Messgerät kontaktieren Sie bitte die PCE Deutschland Gm...

Отзывы: