Manual
7
4.3.2
Perform measurements
Press the Start button when all preparations are made.
The display shows “- - - -“ while the measurement
is made. After the selected area has been detected the probe only moves backwards to the position from
where measurements are started. After this point has been reached, the measured value is displayed on
the screen. Now it is possible to view the results in Ra or Rz by simply pressing the Parameter button.
4.3.3
Set the measurement length
The measurement length is set by making use of the Calibration button. Press the Calibration button until
the display shows “LEN”. You can now choose length “1” and “2” by pressing the Up or Down button. The
settings can be saved by pressing the Start button.
4.3.4
Calibration
To calibrate the device, press the Calibration button.
“CAL” will show up in the display. Now place the
device on the calibration plate and press the Start button. The display will show a result after
measurement has been made. You can now adjust value according to the reference value on the
calibration plate. To do so, use the Up / Down button. When the correct value has been set, switch the
device off and on again.
4.3.5
Set Cutoff length
To set the cutoff length the Cutoff button simply needs to be pressed. You can choose between 2,5 mm,
0,25 mm and 0,8 mm.
4.4
Notes
4.4.1
General Notes
???
???
4.4.2
Standards
-
ISO 4287
-
DIN 4768
-
JIS B601
-
ANSI B46.1