12 

5.2.2 

 Specify measurement / stop mode 

 

After the selection of the heat-up mode, the bar icon will flash in the upper side of the display. 

 

 

 

The user can select between three measurement / stop modes. 

 
 

 

Constant measurement value

 

In this measurement / stop mode, the measurement is stopped automatically when the 

measurement value is constant over a certain period of time. 

The following table shows the stop conditions: 

 

 

Stop conditions 

<period of time (weight change)> 

 

Fast 

heat-up mode 

Standard 

heat-up mode 

Slow 

Heat-up mode 

PCE-MA 110 

36 s (<20 mg) 

45 s (<20 mg) 

60 s (<20 mg) 

PCE-MA 202 

36 s (<20 mg) 

45 s (<20 mg) 

60 s (<20 mg) 

PCE-MA 100 

36 s (<3 mg) 

45 s (<2 mg) 

60 s (<2 mg) 

PCE-MA 200 

36 s (<3 mg) 

45 s (<2 mg) 

60 s (<2 mg) 

 

 

ATTENTION: 

In this mode, it is not possible to set a measuring time. 

 

 

 

Manual / user stop mode 

In this measurement / stop mode, the measurement can be ended by the user. To do so, press 
the 

 key. 

 

ATTENTION:

 Depending on the temperature,

 

combustion of the sample can easily occur in this 

measurement / stop ,mode. This mode should only be used if constant monitoring can be 

guaranteed. The advantage of this mode is that the sample can be dried for a long time to allow 

moisture to escape which is located very deep in the sample. 

 

 

 

Time stop mode

 

In this measurement / stop mode, the measurement is stopped automatically when the set 
measuring / drying time has passed. 

 

 

ATTENTION: 

The maximum measuring / drying time you can set is 99 minutes. 

 

The desired measurement / stop mode can be selected by means of the arrow keys 

 and 

To confirm the selected measurement / stop mode, press the 

 key. 

 

 

Содержание PCE-MA 100

Страница 1: ...8 Jupiter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint B...

Страница 2: ...ion 9 4 3 Disposable aluminium sample pan 9 4 4 Spreading of the sample 9 4 5 Glass fibre filter circle 10 4 6 Practical advice 10 5 Operation 10 5 1 Measurement 10 5 2 Drying modes 11 5 2 1 Specify h...

Страница 3: ...E Instruments or equivalent Before each use inspect the case for visible damage If any damage is visible do not use the device Do not use the instrument in explosive atmospheres The limit values for t...

Страница 4: ...e Measurement range moisture 0 100 Readability moisture 0 01 0 1 0 01 0 1 Memory 16 drying programmes parameters Interface RS 232 Display LCD 17 mm digit height Power supply 220 V 50 Hz or 110 V 60 Hz...

Страница 5: ...tection plate 1 2 The notch in the side wall of the wind shield 2 must be placed in a way that it faces the user 3 The handle of the insertion aid 3 is placed to rest in this notch 4 Place the sample...

Страница 6: ...eat up mode Standard Quick Slow Measuring stop mode constant reading user time lapse Drying temperature 40 C bis zu 199 C Drying time Duration of drying process Percentage moisture Percentage dry weig...

Страница 7: ...on off Start stop measurement Tare Leave settings Open settings Switch between absolute and relative measuring result Confirm current settings go to next setting Return to weighing mode short keystro...

Страница 8: ...optional PCE BP 1 printer to the scales To do so you have to set the following on the printer The printer comes with an RS 232 cable which is used for connecting both devices SUB D 9 plug for RS 232 i...

Страница 9: ...the measurement You should avoid tools with heat conducting properties i e which are able to transfer heat to the sample Improper handling and preparation of the sample will falsify the end result of...

Страница 10: ...ure balance saves the drying parameters last used after switching off and back on These are shown in the display Prepare the sample and place it in the drying chamber You can now start the measurement...

Страница 11: ...an select one out of three heat up speeds Standard heat up mode This is the default mode which is suitable for most sample types In this heat up mode 120 C are reached after approx 4 minutes Quick hea...

Страница 12: ...s 2 mg 60 s 2 mg ATTENTION In this mode it is not possible to set a measuring time Manual user stop mode In this measurement stop mode the measurement can be ended by the user To do so press the key...

Страница 13: ...mperature regulation temperature fluctuations of approx 4 C can occur Mind this when selecting the temperature to make sure the sample does not scorch Otherwise the moisture reading will be falsified...

Страница 14: ...n be adjusted by means of a 100 g external test weight min M1 Special care must be taken when positioning the scales as little vibration as possible and balanced by means of the level The adjustment o...

Страница 15: ...appear in the display which represent the last 5 measurements which are stored in the device Select a measurement by using the arrow keys and and press the key to confirm 5 The data is now sent to th...

Страница 16: ...with the EU directive 2012 19 EU we take our devices back We either re use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please cont...

Страница 17: ...la pr sente d claration est conforme aux normes cit es ci apr s TR Uygunluk Beyan Bu bildirime ba l r n n a a daki standartlara uygun oldu unu beyan ederiz NL Conformiteit verklaring Wij verklaren hi...

Отзывы: