© PCE Instruments
19
Eng
li
s
h
6
Adjustment
For the adjustment of the scales, weights of 5 kg and 15 kg are required. To make an
adjustment, unscrew the scales. The main circuit board has three pins. Pins no. 2 and 3 must
be linked when starting the scales. The link between the pins can then be removed and the
scales can be closed again.
After startup, the display should look as follows:
Now press the key
in unloaded condition.
Display 2 shows
Steps
“1“
Press
→ T ← key
“2“
Apply 5 kg and confirm with
→ T ← key
“3“
Keep 5 kg
on the scales, enter “5000“ in display 1
and confirm with
→ T ← key
“4“
In addition to the 5 kg, add another 10 kg and enter
“30000“ and confirm with the → T ← key
“5“
Display 1 will now show
“3“. This means that the
scales work with 3 decimal places. Confirm with the
→ T ← key
“6“
Display 1 now shows
“10“. This means that the scales
have a resolution of 10. Confirm with
the → T ← key
When the adjustment is completed, the scales will enter normal weighing mode. The accuracy of
the scales can now be checked, for example with a weight of 5 kg or 15 kg.