6.8 

Relative value  

The  relative  value  is  the  difference  in  value  between  the  current  light  value  and  the  saved 
reference light value. 
To save a reading as a reference light value, press the REL key when the desired reference value 
is displayed on the LCD. The value displayed thereafter will be the relative value of that  saved 
value. 
For example: with the saved reference value 80 lux, if the current light value is 100 lux, the meter 
will display 20 lux. To retrieve the reference value, press the REL  key again. The value will be 
displayed and the REL icon will flash on the LCD.  
To exit REL mode, press and hold the REL key for approximately 2 seconds until the REL icon 
on the LCD disappears. 

6.9 

Backlight  

If no key is pressed for 60 seconds in APO (auto power off) mode or recording mode, the backlight 
will automatically dim to save battery power. Press any key to increase the brightness again. In 
non-APO mode, the backlight is always very bright.  

6.10 

Recording mode (PCE-LMD 10 only)  

The lux meter has a data logging function. Before starting the recording function, connect the light 
meter to the PC via Micro-USB and set up the parameters via the Smart Logger software. Set the 
start mode to "by key" and disconnect the light meter from the PC after setup. Press the REC key 
on the lux meter to start recording. The REC icon will start flashing and will be displayed on the 
LCD. 
Do not turn off the meter during recording mode. When the set number of data is reached, the lux 
meter  will  be  turned  off  automatically.  When  the  memory  is  full,  the  FULL  icon  appears  at  the 
bottom of the LCD. When enough data have been deleted, the FULL icon disappears. 
 

Notice

If the  lux  meter  is  turned  off  during  recording, connect  it  to the PC and  set  up  the  parameters 
again after turning it on before starting recording again. Otherwise, the ERR icon will be displayed 
on the LCD when you press the REC key for recording.  

6.11 

Auto Power Off  

The APO function is set to ON by default. To turn off the APO function, press the 

 key lightly. 

To extend the battery life, the lux meter will automatically turn off after about 10 minutes when not 
in use. 
In  record  mode  or  when  the  meter  is  connected  via  USB,  the  APO  function  is  automatically 
disabled until the memory is full or the set number of records is reached. 
 

 

Содержание LMD 10

Страница 1: ...LMD 10 Lux Meter User Manual User manuals in various languages fran ais italiano espa ol portugu s nederlands t rk polski can be found by using our product search on www pce instruments com Last chang...

Страница 2: ...perating instructions 4 6 1 Sensor 4 6 2 Switch on off the meter 4 6 3 Select units 4 6 4 Measurement 4 6 5 Max Min mode 5 6 6 Hold data 5 6 7 Hold a PEAK value 5 6 8 Relative value 6 6 9 Backlight 6...

Страница 3: ...ument when your hands are wet You must not make any technical changes to the device The appliance should only be cleaned with a damp cloth Use only pH neutral cleaner no abrasives or solvents The devi...

Страница 4: ...spectral filter Peak Hold Detection of the peak light intensity of the shortest 10 S Battery level indicator flashes when the battery voltage is too low Data logger PCE LMD 10 32 000 sets of measured...

Страница 5: ...N key PCE LMD 5 MIN MAX key PCE LMD 10 10 MAX key PCE LMD 5 REC key PCE LMD 10 only 11 Light sensor Notice The micro USB socket is located at the bottom of the meter 5 2 Display 1 MAX MIN value 2 Rela...

Страница 6: ...in use 6 2 Switch on off the meter Press and release the key to turn on the light meter and press and hold the same key for 3 seconds to turn off the light meter 6 3 Select units Press the key to sel...

Страница 7: ...rs from the LCD 6 6 Hold data Press the HOLD key The digital reading is held and the HOLD icon appears on the LCD Press the HOLD key again to return to normal operation Note When the Data Hold functio...

Страница 8: ...starting the recording function connect the light meter to the PC via Micro USB and set up the parameters via the Smart Logger software Set the start mode to by key and disconnect the light meter fro...

Страница 9: ...flashes Use the DC 5V power adaptor to connect to the micro USB charging port on the bottom of the meter The battery icon on the LCD will indicate that the battery is charging and will disappear when...

Страница 10: ...Unit 11 Southpoint Business Park Via Pesciatina 878 B Interno 6 Halkal Merkez Mah Ensign Way Southampton 55010 Loc Gragnano Pehlivan Sok No 6 C Hampshire Capannori Lucca 34303 K k ekmece stanbul Unite...

Отзывы: