Manual 

 

6  Other functions 

6.1  Data Hold 

During the measurement, press the " Hold Button " (3-3) once will hold the measured value & the LCD will 
display a " HOLD " symbol. Press the " Hold Button " once again will release the data hold function. 

6.2  Data record (MAX / MIN reading) 

1) The data record function records the maximum and minimum readings. Press the " REC Button " (3-4) 
    once to start the Data Record function and there will be a " REC. " symbol on the display. 
2) With the " REC. " symbol on the display : 
    a) Press the " REC Button " (3-4) once, the" REC. MAX. " symbol along with the maximum value will 
        appear on the display. If you intend to delete the maximum value, just press the " Hold Button " (3-3) 
        once, then the display will show the " REC. " symbol only & execute the memory function  
        continuously. 
    b) Press the " REC Button " (3-4) again, the " REC. MIN. " symbol along with the minimum value will 
    appear on the display. If you intend to delete the minimum value, just press the " Hold Button " (3-3) 
    once, then the display will show the " REC. " symbol only & execute the memory function continuously. 
    c) To exit the memory record function, just press the" REC " button for 2 seconds at least. The display 
        will revert to the current reading. 

6.3  LCD backlight on/off 

After power ON, the " LCD Backlight " will light automatically. During the measurement, press the " 
Backlight Button " (3-2) once will turn OFF the" LCD Backlight ". Press the " Backlight Button " once again 
will turn ON the " LCD Backlight " again. 

7  Datalogger 

7.1  Preparation before using the datalogger function 

a. Insert the SD card 
    Prepare a " SD memory card " ( 1 GB to 16 GB, optional ), insert the SD card into the " SD card 
    socket" (3-20). The front panel of the SD card should face against the down case. 
b. SD card Format 
    If SD card just the first time use into the meter, it recommend to make the " SD card Format " at first. , 
    please see further down. 
c. Time setting 
    If the meter is used at first time, it should to adjust the clock time exactly, please see further down. 
d. Decimal format setting 

 

The numerical data structure of SD card is default used the " . " as the decimal, for 
example "20.6" "1000.53" . But in certain countries ( Europe ...) is used the " , " as the 
decimal point, for example " 20, 6 ", "1000,53". Under such situation, it should change 
the Decimal character at first, details of setting the Decimal point can be found further 
down. 
 

7.2  Auto datalogger (set sampling time

 

 1 second) 

a. Start the datalogger 
    Press the " REC Button (3-4) once , the LCD will show the text " REC ", then press the " Logger 
    Button" (3-8), the bottom text " DATALOGGER " will flash, at the same time the measuring data along 
    the time information will be saved into the memory circuit. 

 

Note: 

* How to set the sampling time, refer to the description further down. 

* How to set the beeper sound is enable, refer to the description further down. 

 
b. Pause the datalogger 
    During execute the Datalogger function , if press the " Logger Button " (3-8) once will pause the 
    Datalogger function ( stop to save the measuring data into the memory circuit temporally ). At the same 
    time the the text of " DATALOGGER " will be no flashing. 
 

Содержание 4250348702241

Страница 1: ...upiter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Busi...

Страница 2: ...time 0 second 10 7 4 How to check the time information 10 7 5 How to check the sampling time information 10 7 6 SD card data structure 11 8 Saving data from the SD card to the computer EXCEL software...

Страница 3: ...anual power off Data hold record max and min reading Microcomputer circuit high accuracy Power by UM3 AA 1 5 V x 6 batteries or DC 9V adapter RS232 USB PC COMPUTER interface Wide applications water co...

Страница 4: ...adjustable from 1 second up to 8 hours 59 min 59 sec Display LCD display 52 x 38 mm Memory SD card from 1 16 GB 2 GB memory card included Interface RS 232 Software optional Power supply 6 x 1 5 V AA b...

Страница 5: ...ton 3 6 Button Function Button 3 7 Time Button 3 8 Logger Button SET Button Sampling check 3 9 Stand 3 10 Battery Compartment Cover 3 11 Battery Cover Screw 3 12 Tripod Fix Nut 3 13 Temp Socket pH ATC...

Страница 6: ...pH The temperature unit is set to Manual ATC without connect the ATC probe Auto power off The sampling time of data logger function is 2 seconds If the meter is connected to the pH electrode for the...

Страница 7: ...unit pH will change to mV Press the Function Button once will return to pH function 2 Prepare the ORP Electrode optional ORP 14 install the Probe Plug of ORP electrode into the pH Socket BNC Socket 3...

Страница 8: ...b screen should cooperate the pH 7 00 standard solution a screen should cooperate the pH 4 00 standard solution Pressing the Enter Button 3 4 will save the calibration data and finish the calibration...

Страница 9: ...CD Backlight again 7 Datalogger 7 1 Preparation before using the datalogger function a Insert the SD card Prepare a SD memory card 1 GB to 16 GB optional insert the SD card into the SD card socket 3 2...

Страница 10: ...ion no will flash It can use the Button 3 5 or Button 3 6 to set the measuring position 1 to 99 for example room 1 to room 99 to identify the measurement location the lower Display will show P x x 1 t...

Страница 11: ...HA02001 XLS PHAA02002 XLS PHA02099 XLS PHAXX Note XX Max value is 10 8 Saving data from the SD card to the computer EXCEL software 1 After executing the Data Logger function take away the SD card out...

Страница 12: ...the display will show Sd F SD memory card Format dAtE Set clock time Year Month Date Hour Minute Second SP t Set sampling time Hour Minute Second PoFF A uto power OFF management bEEP S et beeper soun...

Страница 13: ...he SET Button 3 8 once will save the time value then the screen will jump to Sampling time setting screen see below Note After the time value is setting the internal clock will run precisely even Powe...

Страница 14: ...Button 3 4 will save the setting function with default 9 7 Select the temperature unit C or F When the lower display shows t CF 1 Use the Button 3 5 or Button 3 6 to select the upper Display text to...

Страница 15: ...ce with DC 1 5 V battery UM3 AA Alkaline heavy duty x 6 PCs and reinstate the cover 4 Make sure the battery cover is secured after changing batteries 12 System reset If troubles arise such as CPU syst...

Страница 16: ...or for Display PH 05 mV 18 D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D1 LSD D8 MSD For example If the...

Страница 17: ...use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 15 Contact If you have any questions about our rang...

Отзывы: