1  Introduction 

Thank you for purchasing a PCE-228 from PCE Instruments. 
* pH : 0 to 14.00 pH, mV ( ORP ) : ± 1999 mV. 
* pH measurement can select ATC or manual Temp. adj.. 
* Optional ATC probe for pH measurement. 
* pH measurement can make the auto calibration for pH 7, pH 4 and pH 10 or other value. 
* Real time SD memory card Datalogger, it Built-in Clock and Calendar, sampling time set from 1 sec to 8 
  hour 59 min. 59 sec. 
* Manual datalogger is available ( set the sampling time to 0 ), during execute the manual datalogger 
  function, it can set the different position ( location ) No. ( position 1 to position 99 ). 
* Innovation and easy operation, computer is not need to setup extra software, after execute datalogger, 
  just take away the SD card from the meter and plug in the SD card into the computer, it can down load 
  the all the measured value with the time information ( year/month/date/hour/minute/second ) to the Excel 
  directly, then user can make the further data or graphic analysis by themselves. 
* SD card capacity : 1 GB to 16 GB. 
* LCD with green light backlight, easy reading. 
* Can default auto power off or manual power off. 
* Data hold, record max. and min. reading. 
* Microcomputer circuit, high accuracy. 
* Power by UM3/AA ( 1.5 V ) x 6 batteries or DC 9V adapter. 
* RS232/USB PC COMPUTER interface. 
* Wide applications: water conditioning, aquariums, beverage, fish hatcheries, food processing, 
  photography,laboratory, paper industry, plating industry, quality control, school & college, water 
  conditioning. 

2  Safety notes 

Please read this manual carefully and completely before you use the device for the first time. The device 
may only be used by qualified personnel and repaired by PCE Instruments personnel. There is no 
warranty of damages or injuries caused by non-observance of the manual.

 

 
- The device may only be used in the approved temperature range. 

The case should only be opened by qualified personnel of PCE Instruments.

 

- The instrument should never be placed with the user interface facing an object (e.g. keyboard side on a 
  table). 
- You must not make any technical changes to the device. 
- The appliance should only be cleaned with a damp cloth / use only pH-neutral cleaner.

 

 
This user's handbook is published by PCE Instruments without any guarantee. 
 
We expressly point to our general guarantee terms which can be found in our general terms of business. 
 
If you have any questions please contact PCE Instruments. 
 

 

Содержание 4250348702241

Страница 1: ...upiter FL 33458 USA From outside US 1 Tel 561 320 9162 Fax 561 320 9176 info pce americas com www pce instruments com english www pce instruments com PCE Instruments UK Ltd Units 12 13 Southpoint Busi...

Страница 2: ...time 0 second 10 7 4 How to check the time information 10 7 5 How to check the sampling time information 10 7 6 SD card data structure 11 8 Saving data from the SD card to the computer EXCEL software...

Страница 3: ...anual power off Data hold record max and min reading Microcomputer circuit high accuracy Power by UM3 AA 1 5 V x 6 batteries or DC 9V adapter RS232 USB PC COMPUTER interface Wide applications water co...

Страница 4: ...adjustable from 1 second up to 8 hours 59 min 59 sec Display LCD display 52 x 38 mm Memory SD card from 1 16 GB 2 GB memory card included Interface RS 232 Software optional Power supply 6 x 1 5 V AA b...

Страница 5: ...ton 3 6 Button Function Button 3 7 Time Button 3 8 Logger Button SET Button Sampling check 3 9 Stand 3 10 Battery Compartment Cover 3 11 Battery Cover Screw 3 12 Tripod Fix Nut 3 13 Temp Socket pH ATC...

Страница 6: ...pH The temperature unit is set to Manual ATC without connect the ATC probe Auto power off The sampling time of data logger function is 2 seconds If the meter is connected to the pH electrode for the...

Страница 7: ...unit pH will change to mV Press the Function Button once will return to pH function 2 Prepare the ORP Electrode optional ORP 14 install the Probe Plug of ORP electrode into the pH Socket BNC Socket 3...

Страница 8: ...b screen should cooperate the pH 7 00 standard solution a screen should cooperate the pH 4 00 standard solution Pressing the Enter Button 3 4 will save the calibration data and finish the calibration...

Страница 9: ...CD Backlight again 7 Datalogger 7 1 Preparation before using the datalogger function a Insert the SD card Prepare a SD memory card 1 GB to 16 GB optional insert the SD card into the SD card socket 3 2...

Страница 10: ...ion no will flash It can use the Button 3 5 or Button 3 6 to set the measuring position 1 to 99 for example room 1 to room 99 to identify the measurement location the lower Display will show P x x 1 t...

Страница 11: ...HA02001 XLS PHAA02002 XLS PHA02099 XLS PHAXX Note XX Max value is 10 8 Saving data from the SD card to the computer EXCEL software 1 After executing the Data Logger function take away the SD card out...

Страница 12: ...the display will show Sd F SD memory card Format dAtE Set clock time Year Month Date Hour Minute Second SP t Set sampling time Hour Minute Second PoFF A uto power OFF management bEEP S et beeper soun...

Страница 13: ...he SET Button 3 8 once will save the time value then the screen will jump to Sampling time setting screen see below Note After the time value is setting the internal clock will run precisely even Powe...

Страница 14: ...Button 3 4 will save the setting function with default 9 7 Select the temperature unit C or F When the lower display shows t CF 1 Use the Button 3 5 or Button 3 6 to select the upper Display text to...

Страница 15: ...ce with DC 1 5 V battery UM3 AA Alkaline heavy duty x 6 PCs and reinstate the cover 4 Make sure the battery cover is secured after changing batteries 12 System reset If troubles arise such as CPU syst...

Страница 16: ...or for Display PH 05 mV 18 D10 Polarity 0 Positive 1 Negative D9 Decimal Point DP position from right to the left 0 No DP 1 1 DP 2 2 DP 3 3 DP D8 to D1 Display reading D1 LSD D8 MSD For example If the...

Страница 17: ...use them or give them to a recycling company which disposes of the devices in line with law If you have any questions please contact PCE Instruments 15 Contact If you have any questions about our rang...

Отзывы: