© PCE Instruments
36
5.4
Data logger
The internal memory of the device can save up to 1,800 values. (Each item selected in M50 is
considered a value). To set or start the logger, do the follow ing:
(1) Via w indow M51, you can set the start time, the saving interval and the run time. This means
that measurements can be taken for a maximum of 24 hours.
(2) Window M52 is used to determine w here data are saved. According to the standard setting,
the data are saved to the device. You can also transfer the data directly to the RS-232 / USB
interface w ithout saving them to the internal memory.
(3) Via w indow M50, you can set the logger and select the values you w ould like to save.
You can view the saved data via w indow M53.
You can read out the saved data from the buffer via the RS-232 / USB interface and delete the internal
memory using the functions in w indow M52.
To transfer data to a computer, you w ill need the softw are SOFT-PCE-TDS.
6
Calibration
There is a (calibration) factor betw een the real flow velocity and the flow velocity displayed by the
device. This calibration factor can be determined by carrying out a calibration. How ever, you will need
flow calibration equipment to do so.
Please send the device to PCE Instruments for calibration. Our contact details can be found at the end
of this manual.
7
Maintenance
7.1
Troubleshooting
Error m essages after switching on
The device runs a self-diagnostic programme w hen sw itched on. This diagnostic programme is
supposed to find hardw are errors. The follow ing chart shows possible error messages.
Error m essage
Reason
Countermeasure
„ROM Testing Error”
„Segment Test Error”
Softw are problem
(1) Restart the device
(2) Contact PCE Instruments
„Stored Data Error“
The parameters entered by
the user are not integrated
Press the ENTER key. All
values are reset to default.
„Timer Slow Error”
„Timer Fast Error”
Problems w ith the timekeeper
(1) Restart the device
(2) Contact PCE Instruments
„Date Time Error”
Number error in the calendar
Initialise the calendar via
w indow M61
Repeated reboot
Hardw are problem
Contact PCE Instruments