Model 1102-01A

Strain Gage Load Cell

Installation and Operating Manual

For assistance with the operation of this product,contact: 

PCB Load & Torque, Inc.

Toll-free: 866-684-7107

24-hour SensorLine™: 716-684-0001

Fax: 248-888-8266

E-mail: [email protected] 

Web: www.pcbLoadTorque.com

Содержание 1102-01A

Страница 1: ...nstallation and Operating Manual For assistance with the operation of this product contact PCB Load Torque Inc Toll free 866 684 7107 24 hour SensorLine 716 684 0001 Fax 248 888 8266 E mail LTInfo pcbloadtorque com Web www pcbLoadTorque com ...

Страница 2: ... repair User servicing or repair is not recommended and if attempted may void the factory warranty Calibration Routine calibration of sensors and associated instrumentation is recommended as this helps build confidence in measurement accuracy and acquired data Equipment calibration cycles are typically established by the users own quality regimen When in doubt about a calibration cycle a good rule...

Страница 3: ...ory customer service representative Warranty All equipment and repair services provided by PCB Piezotronics Inc are covered by a limited warranty against defective material and workmanship for a period of one year from date of original purchase Contact PCB for a complete statement of our warranty Expendable items such as batteries and mounting hardware are not covered by warranty Mechanical damage...

Страница 4: ... O PCB板 X O O O O O 电气连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合金 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在 GB T 26572 规定的限量要求以下 X 表示该有害物质至少在该部件的某一均质材料中的含量超出 GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...te O O X O O O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at le...

Страница 6: ...ing Considerations 3 6 0 CALIBRATION 3 6 1 Calibration Certificate Description 3 6 1 1 Measured Output 4 6 1 2 Hysteresis 4 6 1 3 Best Fit Output 4 6 1 4 Strain Gage Measurements 4 6 1 5 Shunt Calibration Standard Resistor 4 6 1 6 Static Error Band SEB 4 7 0 SHUNT CALIBRATION DESCRIPTION 4 7 1 Resistor Value 4 7 2 Shunt Calibration Process 4 7 3 Estimating Shunt Resistor for a Given Load 4 8 0 MAI...

Страница 7: ...nts 4 0 MECHANICAL INSTALLATION Loads applied to canister load cells must be even and centered Compensation linkages e g rod end bearing alignment couplers are recommended to help minimize load misalignment When installing fixtures be sure to thread the fixture to the load cell and not the load cell to the fixture If alignment fixtures are not used ensure loading surfaces are flat and parallel wit...

Страница 8: ...instrument or control system is grounded through its power cord Ground loops and measuring system wiring may result in unstable or noisy signals A simple test with a voltmeter connected between the power cord ground and the structure on which the load cell is mounted can confirm that the structure has been properly grounded If the power cord ground and structure ground are not at the same potentia...

Страница 9: ...leg of the bridge to apply the shunt resistor to for both tension and compression load simulation Typically tension is simulated by inserting the shunt resistor between the P and S connector leads Compression loading is simulated by inserting the shunt resistor between the S and P connector leads 7 1 Resistor Value Canister load cells have a nominal 2 0 mV V full scale output For a 700 ohm strain ...

Страница 10: ...e signal leads 9 Check bridge balance 10 Keep a record of your observations correct problems or contact PCB factory for assistance 10 0 CALIBRATION REPAIR SERVICES PCB Load Torque offers calibration and repair services The PCB Calibration Laboratory in Farmington Hills Michigan is A2LA Accredited per ISO IEC 17025 Standard calibration certificates list five force points ascending and one point des...

Страница 11: ...x 38 1 mm 7 Weight 0 67 lb 304 g Mounting Thread 1 4 28 Female No Metric Equivalent Housing Material Painted Aluminum Painted Aluminum Sensing Element Strain Gage Strain Gage Deflection at Full Scale Capacity 0 003 in 0 08 mm Electrical Connector PT02E 10 6P PT02E 10 6P All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve ...

Страница 12: ...ULL 1102 SERIES PT CONN CANNISTER LOAD CELL PTE 9 14 11 DECIMALS XX 01 XXX 005 DECIMALS X 0 3 XX 0 13 51277 PCB Load Torque Inc claims proprietary rights in the information disclosed hereon Neither it nor any reproduction thereof will be disclosed to others without the written consent of PCB Load and Torque Inc REVISIONS REV DESCRIPTION ECO A ADDED NOTE 1 WIRING CODE 38165 PT02E 10 6P RECEPTACLE T...