Introduction

Model 394C06 Handheld Shaker is a small, handy, completely self-contained vibration reference source. It is

intended for rapid checking of vibration measurement, monitoring and recording systems using piezoelectric

accelerometers, as well as other types of vibration transducers having a maximum weight of 210 grams.

Figure 1: Model 394C06 Handheld Shaker

Description

Model 394C06 permits accurate adjustment of measuring instrumentation to output a standard acceleration level

of 1 g RMS or 1 g peak, with automatic internal compensation for the applied mass. The reference signal may

also be used for velocity and displacement, at 10 mms

-1

RMS or 10 mms

-1

peak and 10 µm RMS or 10 µm peak-

to-peak respectively. To change between RMS and peak, depress the corresponding side of the rocker switch in

the options cavity of the shaker. For example, to choose RMS, depress the left side of the rocker switch as

illustrated in

Figure 1

.

Model 394C06 also provides quick, single sensor validation in complex measurement systems. Confirmation of

proper function through the full measurement chain from the sensor, cable(s), signal conditioning through to the

data acquisition instrumentation, is extremely useful when configuring large sensor arrays.

Low Power

Indicator Light

Power On/Off

(momentary button -

push to apply)

Accelerometer
mounting hole

Options cavity
under

Battery cover
threaded for easy
removal
(4AA batteries)

DC 12 VOLTS/1.6 AMP

Содержание M394C06

Страница 1: ... Hz for up to 210 grams total weight of Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E mail info pcb com Web www pcb com ...

Страница 2: ...zed tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak tes ng hydrosta c pressure tes ng and others For more informa on contact your local PCB Piezotronics distributor sales representa ve or factory customer service representative Returning Equipment If factory repair is required our representatives will prov...

Страница 3: ...ma on on par cular opera ng steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety precau ons to avoid personal contact with this voltage This symbol on the unit indicates that the user should refer to the opera ng instruc ons located in the manual This symbol indicates safety earth g...

Страница 4: ...BDE 住房 O O O O O O PCB板 X O O O O O 电 连接器 O O O O O O 压电晶体 X O O O O O 环氧 O O O O O O 铁氟龙 O O O O O O 电子 O O O O O O 厚膜基板 O O X O O O 电线 O O O O O O 电缆 X O O O O O 塑料 O O O O O O 焊接 X O O O O O 铜合 黄铜 X O O O O O 本表格依据 SJ T 11364 的规定编制 O 表示该有害物质在该部件所有均质材料中的含量均在GB T 26572 规定的限量要求以下 X 表示该有害物质 少在该部件的某 均质材料中的含量超出GB T 26572 规定的限量要求 铅是欧洲RoHS指令2011 65 EU附件三和附件四目前由于允许的豁免 CHINA RoHS COMPLIANCE ...

Страница 5: ...O O Wires O O O O O O Cables X O O O O O Plas c O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that said hazardous substance contained in all of the homogeneous materials for this part is below the limit requirement of GB T 26572 X Indicates that said hazardous substance contained in at least one of t...

Страница 6: ...ndheld Vibration Shaker Operating Guide with Enclosed Warranty Information 3425 Walden Avenue Depew New York 14043 2495 Phone 716 684 0003 Fax 716 684 3823 Toll Free Line 1 800 828 8840 MANUAL NUMBER 18295 MANUAL REVISION B ECO 51616 ...

Страница 7: ...MBER 18295 MANUAL REVISION B Table of Contents Introduction page 3 Description page 3 Operation page 4 Warnings Cautions page 6 Warranty page 7 Service page 7 Return Procedure page 7 Total Customer Satisfaction page 8 ...

Страница 8: ...also be used for velocity and displacement at 10 mms 1 RMS or 10 mms 1 peak and 10 µm RMS or 10 µm peak to peak respectively To change between RMS and peak depress the corresponding side of the rocker switch in the options cavity of the shaker For example to choose RMS depress the left side of the rocker switch as illustrated in Figure 1 Model 394C06 also provides quick single sensor validation in...

Страница 9: ...gure 1 The unit will turn on and internal power to the shaker will be ramped up providing a smooth gradual increase in vibration amplitude This process prevents shock induced overloads to the test sensor A special reference sensor within the unit constantly measures the g level and internal controls adjust the shaker magnitude to maintain 1 g output while maintaining the frequency of 159 2 Hz A re...

Страница 10: ...Points Calibration of the internal feedback circuit can be verified via test points located in the options cavity see Figure 1 The measured value in volts AC should agree with the calibration certificate This is useful for a quick field check of unit integrity which may be necessary if the shaker is dropped and a reference standard is not immediately available This test is not a guarantee of shake...

Страница 11: ...rge ESD can damage electrical devices Similar to a capacitor a cable can hold a charge caused by triboelectric transfer such as that which occurs in the following Laying on and moving across a rug Any movement through air The action of rolling out a cable and or Contact with a non grounded person The PCB solution for product safety Connect the cables only with the AC power off Temporarily short th...

Страница 12: ...t quotation will be provided prior to servicing at no charge Before returning the unit please consult a factory PCB applications engineer concerning the situation as certain problems can often be corrected with simple on site procedures Return Procedure To expedite returned instrumentation contact a factory PCB applications engineer for a RETURN MATERIAL AUTHORIZATION RMA NUMBER When requesting th...

Страница 13: ...eplace or exchange it at no charge You may also choose within the warranty period to have your purchase price refunded PCB offers to all customers at no charge 24 hour phone support This service makes product or application support available to our customers day or night seven days a week When unforeseen problems or emergency situations arise call the PCB Hot Line at 716 684 0003 and an applicatio...

Страница 14: ... Mounting Torque Maximum 10 in lb 112 N cm 11 All specifications are at room temperature unless otherwise specified In the interest of constant product improvement we reserve the right to change specifications without notice ICP is a registered trademark of PCB Group Inc OPTIONAL VERSIONS Optional versions have identical specifications and accessories as listed for the standard model except where ...

Страница 15: ...MOUNTING HOLE PREPARATION n4 22 n 166 X 7 62 300 x MIN M5 X 0 8 6H X 5 08 200 x MIN E 1 4 28 MOUNTING INSTRUCTIONS METRIC DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTING TORQUE 2 5 FOOT POUNDS 3 7 NEWTON METERS MOUNTING HOLE PREPARATION n 218 n5 54 X 300 7 62 x MIN 1 4 28 UNF 2B X 200 5 08 x MIN F M6 X 0 75 M6 X 1 00 M8 X 1 25 MOUNTING INSTRUCTIONS ENGLISH DIMENSIONS IN BRACKETS 4 RECOMMENDED MOUNTI...

Страница 16: ... GmbH nbn nbn at www nbn at 43 316 40 28 05 Riesstraße 146 8010 Graz Aufgrund laufenderWeiterentwicklungen sind Änderungen der Spezifikationen vorbehalten Alle Angaben vorbehaltlich Satz und Druckfehler v31 05 21 ...

Отзывы: