                 443B101/443B102       

 3 

 

Manual 19854 Rev. D 
ECN 50523 

 

high  temperature  transient  sensitivity  are  used,  the 
high pass filter of the amplifier can be changed from 
0.2  Hz  to  2  Hz,  reducing  the  influence  from 
temperature-induced signals.  The overall gain of the 
amplifier is the ratio of the output sensitivity to the 
sensor sensitivity. 

 
Low Pass Filters 

 

Six selectable upper cutoff frequencies  are provided 
by  the  low  pass  filter,  located  just  after  the  input 
amplifier  to  filter  out  unwanted  signal  components 
before  further  amplification.  The  filters  are  of  the 
two-pole configuration, giving a 12 dB/octave cutoff.  
The low pass filter section may be turned off giving a 
frequency response of >200 kHz.   

 

Integrator Amplifiers 

 

The  integrator  amplifiers  provide  single  integration 
for velocity and double integration for displacement.  
Two low-frequency cutoff settings of 1 or 10 Hz are 
provided in the velocity and displacement modes to 
suppress low-frequency noise.   
 

Output Amplifier 

 

The output  amplifier provides the signal to the con-
tinuous  gain  stage  and  then  to  the  BNC  output 
connector on the front panel. It provides a fixed gain 
of x1 or x10. 
 
Typical  frequency  response  as  a  function  of 
capacitive load of the output of the 443B dual mode 
amplifier  is  shown  in  Figure  7  of  Appendix  B.  Note 
that for a frequency range up to 10kHz, a capacitive 
load  of  up  to  50  nF  may  be  applied.    For  standard 
cable  with  a  shunt  capacitance  of  100pF/m,  up  to 
500  m  of  output  cable  can  be  used  to  cover  this 
frequency range. 

 

Overload Detector 

 

The  overload  detector  monitors  the  output  of  the 
charge amplifier, the low pass filter, and the output, 
so  overload  in  various  parts  of  the  circuitry  are  not 
masked by filtering. The overload indicator is located 
on the front panel.  In ICP® mode, the sensor output 
is also monitored for overloads.

 

Reference Oscillator 

 

The  Model  443B  has  a  built-in  reference  signal 
source, providing a signal of 100 pC RMS at 159.2 Hz 
in charge mode.  In ICP

®

 mode, the reference signal is 

1  V  RMS.    These  signals  can  be  used  as  a  reference 
when using portable DAT recorders. 

Operation 

 

Set-Up and Measurement  

 
1.

 

Connect  the  Model  443B  to  a  suitable  power 
source.  Switch the unit on and let it stabilize for 
30 seconds. 

2.

 

Connect  the  output  of  the  443B  to  a  suitable 
indicating  instrument,  voltmeter,  scope  or 
frequency analyzer. 

3.

 

Select  a  suitable  accelerometer  to  be  used  for 
the  measurement  and  observe  the  correct 
mounting  and  installation  instructions.  Connect 
it to the input connector of the Model 443B. 

4.

 

Refer to the acc

elerometer’s calibration chart to 

select the sensitivity.  

 

Setting the Operational Mode 

 

1.

 

Press the MENU button on the Model 443B. The 
main menu appears on the screen of the Model 
443B. 

2.

 

Select the MODE position with the down button. 

3.

 

Press  the  MENU  button  again.  The  display  now 
shows the following text: 

 

Charge    

Ref. On 

ICP   

Ref. Off 

 

4.

 

Select the mode by moving the cursor to Charge 
or ICP® depending upon which type of sensor is 
being  used.  Note  that  in  ICP®  mode,  zero 
current  can be selected, which  will put the unit 
in voltage mode. 

5.

 

The internal sinusoidal reference may be turned 
on by moving the cursor to Ref. On and pressing 
the  MENU  button.  To  turn  off  the  reference, 
move the cursor to Ref. Off in the MODE menu 
and press MENU. 

 

Setting the Sensor Sensitivity 

 

1.

 

Press the MENU button again and select SEN  

Содержание 441A101

Страница 1: ...01 441A33 443B102 installation and operating manuals that comprise a Model 443B02 MODULAR AC LINE POWER SUPPLY kit For assistance with the operation of this product contact the PCB Piezotronics Inc To...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...WER SUPPLY Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Страница 7: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 8: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 9: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 10: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 11: ...41A43 441A45 AND 441A49 2 EQUIPMENT RATINGS 2 MULTIPLE CHASSIS CONFIGURATION 2 CHANGING THE FUSE 2 DISABLING COOLING FAN 2 AVAILABLE MODULES 2 AC POWER SUPPLY MODEL 441A101 3 DC POWER SUPPLY MODEL 441...

Страница 12: ...tion High voltage electrostatic discharge ESD can damage electrical devices To avoid triboelectric transfer 1 Connect the cables only with the AC power off 2 Temporarily short the end of the cable bef...

Страница 13: ...n 3 0 Ah NiMH rechargeable battery The optional Model 488A09 is a 12VDC 3 4A AC Power Adaptor that allows the 441A102 to be used with any AC line power in the world 100 240VAC 50 60Hz The 441A102 can...

Страница 14: ...CHASSIS AND POWER SUPPLY 4 Manual 19680 Rev C ECN 50523 Appendix A Figures...

Страница 15: ...CHASSIS AND POWER SUPPLY 5 Manual 19680 Rev C ECN 50523...

Страница 16: ...Conditioner Card Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 09...

Страница 17: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 18: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 19: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 20: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 21: ...3 OPERATION 3 SET UP AND MEASUREMENT 3 SETTING THE OPERATIONAL MODE 3 SETTING THE SENSOR SENSITIVITY 3 SETTING THE OUTPUT SENSITIVITY 4 SETTING THE TIME CONSTANT 4 ZEROING THE OUTPUT IN LONG TIME CON...

Страница 22: ...s has its signal ground connected via the chassis to mains ground and b the housing of the transducer is isolated from grounded measurement sources If the measurement set up is mounted in a metal inst...

Страница 23: ...of the charge amplifier the low pass filter and the output so overload in various parts of the circuitry are not masked by filtering The overload indicator is located on the front panel In ICP mode th...

Страница 24: ...is selected in charge mode 443B102 only the display will briefly indicate Time Constant is 10 000 sec or Time Constant is 100 000 sec depending on the gain selected And then Autozero in Progress The...

Страница 25: ...Features The 443B incorporates several features which allow the use of sensors with TEDS electronics The 443B can read the raw data from any TEDS sensor and is able to locally decode the data from sen...

Страница 26: ...surements Using the Setup Memory The 443B has the capability to save and recall up to five different setups so that the user can easily switch between frequently used configurations To save a setup 1...

Страница 27: ...associated data See below for command structure and command set for examples End Byte ETX 0x03 marks the end of the message Data Checksum 16 bits the ASCII hex of the 8 bit sum ignoring overflows of...

Страница 28: ...cating with internal module T Time out message not sent rack slot address may be incorrect NAK Reason Bytes Command format is shown below Module Type Three ASCII printable characters This field identi...

Страница 29: ...t is a common or CMM command Command String XYCMMMMOD Description X Rack number of the target module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printab...

Страница 30: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 CMM Module type for the target Specifying CMM in this argument indicates the intention of any module type that may be...

Страница 31: ...ugh 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 CMM Module type for the target Specifying CMM in this argument indicates the intention of any module type that may...

Страница 32: ...Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 Range 3 ASCII printable characters O...

Страница 33: ...the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 XDCR Commands the...

Страница 34: ...ge ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443...

Страница 35: ...the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 ICPM Commands module to charge mode ZZ specifies the constant current as follows 00 turns the constan...

Страница 36: ...cifying C02 addresses a 443B102 SETF Commands module to change the low pass filter setting Z specifies the low pass filter setting as follows 0 turns the low pass filter off 1 sets the low pass filter...

Страница 37: ...rough 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 LOWF Commands module to change the low frequency response Z specifies the...

Страница 38: ...the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 INTG Commands module to engage integration circuitry Z specifies the integrator as follows 1 engages...

Страница 39: ...rget module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 INTU Commands module to s...

Страница 40: ...able characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifyin...

Страница 41: ...able characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifyin...

Страница 42: ...e ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command si...

Страница 43: ...rough 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have lon...

Страница 44: ...of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have long time constant NULL C...

Страница 45: ...ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command sinc...

Страница 46: ...module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 does not have long time constant OFFS Commands the mo...

Страница 47: ...haracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C02 Module type for the target Only type C02 443B102 can accept this command since the 443B101 doe...

Страница 48: ...s Filter Integration Units Reference Overvoltage Fault ICP mode only Zero Lock long time constant charge mode only Description Mode CHRG Charge mode ICP 0mA ICP mode with 0mA constant current voltage...

Страница 49: ...r set to 3 0 kHz 10 kHz Low pass filter set to 10 kHz 30 kHz Low pass filter set to 30 kHz 100 kHz Low pass filter set to 100 kHz Integration Units SI SI units will be used for input and output Eng En...

Страница 50: ...follows ICP mode with a constant current of 2 mA output sensitivity is 10 00 mV unit transducer sensitivity is 1 023 mV unit low frequency response is 2 0 Hz low pass filter set to 10 kHz integration...

Страница 51: ...Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDR Commands the module to return da...

Страница 52: ...printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 spe...

Страница 53: ...t module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDD Commands the module to...

Страница 54: ...aracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 a...

Страница 55: ...aracters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 a...

Страница 56: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102...

Страница 57: ...3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102...

Страница 58: ...rget module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument addresses a 443B101 specifying C02 addresses a 443B102 TEDW tells the module to...

Страница 59: ...in the application register or application register scratchpad if unlocked of a TEDS device Response String 8 bytes of hex data in ASCII format Example Command String 06C02RDAR Response String AABBCC...

Страница 60: ...xample Command String 06C02WRARAABBCCDDEEFFAABB Response String 0 The 443B102 in rack 0 and slot 6 is commanded to write 8 bytes of hex data 06C02WRARAABBCCDDEEFFAABB to the application register of a...

Страница 61: ...et module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable characters 0 through 7 C0Z Module type for the target Specifying C01 in this argument ad...

Страница 62: ...as been received Command String XYC0ZLKAR Description X Rack number of the target module Range ASCII printable characters 0 through 3 Y Slot number of the target module Range ASCII printable character...

Страница 63: ...443B101 443B102 43 Manual 19854 Rev D ECN 50523 Appendix B Figures...

Страница 64: ...443B101 443B102 44 Manual 19854 Rev D ECN 50523...

Страница 65: ...443B101 443B102 45 Manual 19854 Rev D ECN 50523...

Страница 66: ...443B101 443B102 46 Manual 19854 Rev D ECN 50523...

Страница 67: ...443B101 443B102 47 Manual 19854 Rev D ECN 50523...

Страница 68: ...443B101 443B102 48 Manual 19854 Rev D ECN 50523...

Страница 69: ...10 kHz 0 07 V Hz 143 dB 6 Broadband Electrical Noise 1 Hz to 10 kHz Charge Input 9 V 101 dB 7 8 Broadband Electrical Noise 2 Hz to 22 4 kHz Charge Input 5 fC 5 fC 7 8 Spectral Noise 1 kHz 0 8 V Hz 122...

Страница 70: ...WER SUPPLY Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 0987 E ma...

Страница 71: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 72: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 73: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 74: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 75: ......

Страница 76: ...ol for Series 440 modules Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 71...

Страница 77: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 78: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 79: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 80: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 81: ......

Страница 82: ......

Страница 83: ...Conditioner Card Installation and Operating Manual For assistance with the operation of this product contact PCB Piezotronics Inc Toll free 800 828 8840 24 hour SensorLine 716 684 0001 Fax 716 684 09...

Страница 84: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 85: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 86: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 87: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 88: ...CES IN ENGLISH CABLE TOLERANCES IN METRIC 1 LENGTH 1 1 0 1 LENGTH 5 2 0 5 LENGTH 100 6 0 100 LENGTH 1 0 2 54cm LENGTH 30 5cm 2 54cm 0 30 5cm LENGTH 1 5m 5 1cm 0 1 5m LENGTH 30 5m 15 2cm 0 30 5m LENGTH...

Страница 89: ...01 441A33 443B102 installation and operating manuals that comprise a Model 443B02 MODULAR AC LINE POWER SUPPLY kit For assistance with the operation of this product contact the PCB Piezotronics Inc To...

Страница 90: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 91: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 92: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 93: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 94: ...CES IN ENGLISH CABLE TOLERANCES IN METRIC 1 LENGTH 1 1 0 1 LENGTH 5 2 0 5 LENGTH 100 6 0 100 LENGTH 1 0 2 54cm LENGTH 30 5cm 2 54cm 0 30 5cm LENGTH 1 5m 5 1cm 0 1 5m LENGTH 30 5m 15 2cm 0 30 5m LENGTH...

Отзывы: