OPERATING GUIDE FOR 130 SERIES MICROPHONES 

Contents: 

1.0 Introduction ............................... 1

 

2.0 Product Description .................. 1

 

3.0 Powering ................................... 2

 

4.0 Installation ................................. 3

 

5.0 Calibration ................................. 3

 

6.0 Service ....................................... 7

 

7.0 Warranty .................................... 7 

 

 

 
 
 
 
 
 
 
 
 
 
 
 
          Figure 1 

1.0 Introduction 

PCB

®

 130 series ICP® array microphones provide an extremely cost-effective method for large 

channel count sound pressure measurements. 

Multiple array microphones with their excellent 

phase characteristics can be combined with the appropriate software to perform measurements 
that are not possible with a single microphone.

 Typical applications include sound pressure 

mapping, beamforming, acoustic mode analysis, near-field acoustic holography, or sound 
intensity measurments. To ensure the best performance from your 130 series microphone follow 
the installation and operating procedures in this guide.  If you have additional questions 
concerning the microphone system or applications, call a factory Application Engineer at 716-
684-0001, or your nearest PCB

®

 Field Application Engineer. 

2.0 Product Description 

PCB

®

’s 130 series ICP® Array are integrated systems that include a built-in preamplifier.  

Models 130F20, 130F21, and 130F22 are all ¼-inch microphones that provide a 45 mV/Pa 
output over an extended 120dB dynamic range. They are ideal for use at frequencies from 20 Hz 
to 10 kHz (± 2dB) or 20 kHz (± 5dB).  All models are of rugged, stainless steel construction .  

 

 

 

130A23 

130F20

 

130F21 

130F22 

130A24 

Содержание 130A25

Страница 1: ...EAMPLIFIER 10 32 CONNECTOR TEDS Installation and Operating Manual For assistance with the operation of this product contact the PCB Piezotronics Inc Toll free 716 684 0001 24 hour SensorLine 716 684 0...

Страница 2: ...alized tests including sensitivity at elevated or cryogenic temperatures phase response extended high or low frequency response extended range leak testing hydrostatic pressure testing and others For...

Страница 3: ...tion on particular operating steps The following symbols may be found on the equipment described in this manual This symbol on the unit indicates that high voltage may be present Use standard safety p...

Страница 4: ...China RoHS 2 Disclosure Table Pb Hg Cd Cr VI PBB PBDE O O O O O O PCB X O O O O O O O O O O O X O O O O O O O O O O O O O O O O O O O O O O O O O X O O O O O O O O O X O O O O O O O O O O O X O O O O...

Страница 5: ...O Wires O O O O O O Cables X O O O O O Plastic O O O O O O Solder X O O O O O Copper Alloy Brass X O O O O O This table is prepared in accordance with the provisions of SJ T 11364 O Indicates that sai...

Страница 6: ...oustic mode analysis near field acoustic holography or sound intensity measurments To ensure the best performance from your 130 series microphone follow the installation and operating procedures in th...

Страница 7: ...g distances with negligible signal loss 3 0 Powering All ICP powered microphones require constant current excitation for proper operation For this reason use only PCB constant current signal condition...

Страница 8: ...he end of the housing Sharp edged tools can be used to aid in cover removal but it is important to keep the tool at a shallow angle DO NOT make contact with the microphone s delicate diaphragm right h...

Страница 9: ...ce with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards microphones and acoustic calibrators C...

Страница 10: ...e with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards microphones and acoustic calibrators Ca...

Страница 11: ...ibration is performed in compliance with ISO 9001 ISO 10012 1 and ANSI NCSL Z540 3 4 Due to state of the art limitations 4 1 calibration ratios are not possible on pressure measurement standards micro...

Страница 12: ...ver the microphone output may be affected by harsh environments such as moisture dirt mechanical shock or other unusual conditions This may manifest itself in a number of ways ranging from a loss in f...

Страница 13: ...828 8840 ISO 9001 CERTIFIED AS9100 CERTIFIED In the interest of constant product improvement specifications are subject to change without notice Copyright PCB Group Inc 2016 PCB and ICP are registered...

Страница 14: ...7 dB 1 Electrical Excitation Voltage 18 to30 VDC 18 to30 VDC Constant Current Excitation 2 to20 mA 2 to20 mA Output Bias Voltage 5 5 to14 VDC 5 5 to14 VDC Output Impedance 150 Ohm 150 Ohm Physical Ho...

Страница 15: ...003 005 FILLETS AND RADII 0 07 0 13 OUTLINE DRAWING 72402 1 OF 1 5X MODEL 130A25 ICP MICROPHONE DECIMALS XX 03 XXX 010 DECIMALS X 0 8 XX 0 25 JJF 3 23 20 JJF 3 23 20 MJN 3 23 20 72402 PCB Piezotronic...

Отзывы: