1. Specifications and Part Descriptions

2. Removal/Installation of Bur/U-type Attachment onto Nosecone

Holding the nosecone, twist chuck open/close ring in the direction indicated 

by the black arrow until the chuck clicks open and pull the bur blank out.

Bur

- Insert bur spindle into chuck all the way until the spindle touches the back 

of the chuck.

- Place the nosecone upward so that the inserted bur does not slide down and 

then twist the chuck open/close ring in the direction indicated by the white 

arrow until the chuck clicks shut.

- Make sure that the bur is securely installed onto the nosecone, pulling the 

bur with fingers.

U-type prophy/contra angles

- Insert spindle into chuck all the way until the back-end of the U-type 

attachment contacts the shoulder of the 

nose, aligning the slit at the back-end of 

the U-type attachment with protuberance 

on the nose.

- Twist chuck open/close ring in the 

direction indicated by the white arrow 

until the chuck clicks shut.

Note:

 Maximum speed may vary slightly depending on the type of tubing 

connected to the airmotor. 

Operation Manual

E-type Straight Nosecone

043-1023   EX-6

Caution before use

 • The Patterson E-type straight nosecone is intended for use by dental 

professionals only. 

 • Read this operation manual carefully before using the nosecone. 

 • Be sure to follow all the operating instructions, including all cautions and 

warnings.

 • Keep this Operation Manual for future reference.

 • Consider primarily the safety of the patient and the operator when using 

the nosecone. 

ISO E-type

U-type

Applicable to ISO 2.35mm diameter spindle

1:1 drive

40,000rpm

1.98oz (56g)

Back-end configuration

Nose configuration

Chuck configuration

Speed ratio

Allowance maximum speed

Weight

 • Never install a U-type attachment with a bent spindle or a damaged 

bur, etc., onto the nosecone. Such a spindle or bur may not only injure 

the patient but critically damage the nosecone.

 • Never install a non-ISO standard bur or U-type attachment with a 

non-ISO standard spindle.

 • Never open/close the chuck without installing the instrument onto the 

nosecone, as the chuck retaining intensity may be prematurely 

reduced.

 • Never use the nosecone, if the U-type attachment or bur cannot be 

installed securely. 

In such cases, consult the Patterson Service 

Center. 

3. Installation/Removal of Nosecone from/onto motor and Operation

- Make sure that the motor onto which the nosecone is connected is an ISO 

E-type.

- Insert the back-end of the nosecone into the motor.

- Make sure that the nosecone is securely installed onto the motor by pulling 

it gently with fingers before use.

Caution

 • If the nosecone cannot be smoothly installed/removed onto/from the 

motor or if the nosecone rattles after installation, do not use the 

nosecone. 

 • Stop using the air-motor whenever any abnormality in performance, 

sound level, heat level, etc., is sensed in use.

In such cases, consult the Patterson Service Center

Warning

 • Never dismantle or alter the nosecone.

 • Never run the nosecone above 40,000rpm.

4. Cleaning and Drying

The nosecone must be cleaned and lubricated after each use and before 

autoclaving. 

- Remove the nosecone from the motor. Remove the bur or U-type 

attachment used.

Cleaning and Drying the Exterior

Wash the nosecone under running water following the procedures below. 

(Water condition: ≤100°F, ≥3.5L/min; water should be of the same quality as 

drinking water)

- Clean the external surfaces of the 

nosecone with a soft bristled toothbrush 

for 15 seconds or more. 

- Clean around the bur insertion hole for 15 

seconds or more, using an interdental 

brush (wire thickness 0.7mm) that 

conforms to size 4 in ISO 16409.

Under appropriate lighting (natural office 

lighting, 500 lx or higher), inspect the 

nosecone for blood or other debris. If any 

visible debris remains, repeat the process 

until the nosecone is visually clean.

After cleaning, if the residual moisture is present, wipe off with dry cloth or 

blow it off with compressed air (≤50.75psi) until there is no moisture in the 

interior and exterior.

When blowing off with compressed air, cover the nosecone with cloth to 

prevent scattering of water. Proceed to 5: Cleaning and Lubrication.

5. Cleaning and Lubrication

Cleaning and Lubricating the Interior

- Attach an E-Type spray nozzle to the 

nozzle of the PANA SPRAY Plus.

- Shake the can 3 or 4 times and insert the 

E-Type spray nozzle into the rear of the 

nosecone while holding the head of the 

nosecone with a cloth etc.

- Hold the nosecone and spray for 2 or 3 

seconds until oil comes out of the nosecone tip. Repeat lubrication until 

foreign material stops coming out of the tip. 

6. Sterilization

Autoclave is recommended for sterilization.

- Be sure to lubricate the nosecone before sterilization, following step 5: 

Cleaning and Lubrication. 

- Autoclave 275 ˚F for 20 minutes.

7. Warranty

Patterson guarantees the Patterson nosecone for 10 months subject to the 

observance of maintenance instructions, cautions, warnings and normal use. 

Note: 

- Patterson reserves the right to analyze and determine the cause of any 

failure.

- Guarantee is void if breakage, accident or any other issues are caused by 

neglecting daily maintenance, cautions, warnings, or any other misuse and 

common sense in use.

- Dismantling or reassembling by any third parties other than the Patterson 

Service Center voids guarantee.

• Do not use sprays other than PANA SPRAY Plus. Using sprays other 

than PANA SPRAY Plus may cause the nosecone to overheat.

• Hold the spray can upright.

• Firmly hold the nosecone to prevent it slipping when spray pressure is 

applied.

• Spray lubricant until it expels from the nosecone tip.

Caution

 • Put the nosecone into a sterilization pouch by itself. 

 • Never use dry heat sterilization, cold sterilization, etc. 

 • Never autoclave the nosecone combined with others in the same 

sterilization pouch.

 • Never place the nosecone at the bottom of an autoclave chamber, as 

the temperature may rise above the actual setting temperature.

Specifications

Part descriptions

U-type Nose

Protuberance

E-type back-end

Twist-type chuck open/close ring

Shoulder

: AUTOCLAVABLE

OM-HK129E 001

2021-11-00  CACB1100 N