Close
E-Type spray nozzle
15sec.
Caution
Caution
1. Specifications and Part Descriptions
2. Removal/Installation of Bur/U-type Attachment onto Nosecone
Holding the nosecone, twist chuck open/close ring in the direction indicated
by the black arrow until the chuck clicks open and pull the bur blank out.
Bur
- Insert bur spindle into chuck all the way until the spindle touches the back
of the chuck.
- Place the nosecone upward so that the inserted bur does not slide down and
then twist the chuck open/close ring in the direction indicated by the white
arrow until the chuck clicks shut.
- Make sure that the bur is securely installed onto the nosecone, pulling the
bur with fingers.
U-type prophy/contra angles
- Insert spindle into chuck all the way until the back-end of the U-type
attachment contacts the shoulder of the
nose, aligning the slit at the back-end of
the U-type attachment with protuberance
on the nose.
- Twist chuck open/close ring in the
direction indicated by the white arrow
until the chuck clicks shut.
Note:
Maximum speed may vary slightly depending on the type of tubing
connected to the airmotor.
Operation Manual
E-type Straight Nosecone
043-1023 EX-6
Caution before use
• The Patterson E-type straight nosecone is intended for use by dental
professionals only.
• Read this operation manual carefully before using the nosecone.
• Be sure to follow all the operating instructions, including all cautions and
warnings.
• Keep this Operation Manual for future reference.
• Consider primarily the safety of the patient and the operator when using
the nosecone.
ISO E-type
U-type
Applicable to ISO 2.35mm diameter spindle
1:1 drive
40,000rpm
1.98oz (56g)
Back-end configuration
Nose configuration
Chuck configuration
Speed ratio
Allowance maximum speed
Weight
• Never install a U-type attachment with a bent spindle or a damaged
bur, etc., onto the nosecone. Such a spindle or bur may not only injure
the patient but critically damage the nosecone.
• Never install a non-ISO standard bur or U-type attachment with a
non-ISO standard spindle.
• Never open/close the chuck without installing the instrument onto the
nosecone, as the chuck retaining intensity may be prematurely
reduced.
• Never use the nosecone, if the U-type attachment or bur cannot be
installed securely.
In such cases, consult the Patterson Service
Center.
3. Installation/Removal of Nosecone from/onto motor and Operation
- Make sure that the motor onto which the nosecone is connected is an ISO
E-type.
- Insert the back-end of the nosecone into the motor.
- Make sure that the nosecone is securely installed onto the motor by pulling
it gently with fingers before use.
Caution
• If the nosecone cannot be smoothly installed/removed onto/from the
motor or if the nosecone rattles after installation, do not use the
nosecone.
• Stop using the air-motor whenever any abnormality in performance,
sound level, heat level, etc., is sensed in use.
In such cases, consult the Patterson Service Center
Warning
• Never dismantle or alter the nosecone.
• Never run the nosecone above 40,000rpm.
4. Cleaning and Drying
The nosecone must be cleaned and lubricated after each use and before
autoclaving.
- Remove the nosecone from the motor. Remove the bur or U-type
attachment used.
Cleaning and Drying the Exterior
Wash the nosecone under running water following the procedures below.
(Water condition: ≤100°F, ≥3.5L/min; water should be of the same quality as
drinking water)
- Clean the external surfaces of the
nosecone with a soft bristled toothbrush
for 15 seconds or more.
- Clean around the bur insertion hole for 15
seconds or more, using an interdental
brush (wire thickness 0.7mm) that
conforms to size 4 in ISO 16409.
Under appropriate lighting (natural office
lighting, 500 lx or higher), inspect the
nosecone for blood or other debris. If any
visible debris remains, repeat the process
until the nosecone is visually clean.
After cleaning, if the residual moisture is present, wipe off with dry cloth or
blow it off with compressed air (≤50.75psi) until there is no moisture in the
interior and exterior.
When blowing off with compressed air, cover the nosecone with cloth to
prevent scattering of water. Proceed to 5: Cleaning and Lubrication.
5. Cleaning and Lubrication
Cleaning and Lubricating the Interior
- Attach an E-Type spray nozzle to the
nozzle of the PANA SPRAY Plus.
- Shake the can 3 or 4 times and insert the
E-Type spray nozzle into the rear of the
nosecone while holding the head of the
nosecone with a cloth etc.
- Hold the nosecone and spray for 2 or 3
seconds until oil comes out of the nosecone tip. Repeat lubrication until
foreign material stops coming out of the tip.
6. Sterilization
Autoclave is recommended for sterilization.
- Be sure to lubricate the nosecone before sterilization, following step 5:
Cleaning and Lubrication.
- Autoclave 275 ˚F for 20 minutes.
7. Warranty
Patterson guarantees the Patterson nosecone for 10 months subject to the
observance of maintenance instructions, cautions, warnings and normal use.
Note:
- Patterson reserves the right to analyze and determine the cause of any
failure.
- Guarantee is void if breakage, accident or any other issues are caused by
neglecting daily maintenance, cautions, warnings, or any other misuse and
common sense in use.
- Dismantling or reassembling by any third parties other than the Patterson
Service Center voids guarantee.
• Do not use sprays other than PANA SPRAY Plus. Using sprays other
than PANA SPRAY Plus may cause the nosecone to overheat.
• Hold the spray can upright.
• Firmly hold the nosecone to prevent it slipping when spray pressure is
applied.
• Spray lubricant until it expels from the nosecone tip.
Caution
• Put the nosecone into a sterilization pouch by itself.
• Never use dry heat sterilization, cold sterilization, etc.
• Never autoclave the nosecone combined with others in the same
sterilization pouch.
• Never place the nosecone at the bottom of an autoclave chamber, as
the temperature may rise above the actual setting temperature.
Specifications
Part descriptions
U-type Nose
Protuberance
E-type back-end
Twist-type chuck open/close ring
Shoulder
: AUTOCLAVABLE
OM-HK129E 001
2021-11-00 CACB1100 N