Measuring Particle Size and Zeta Potential
NanoPlus Operator’s Manual
4-10
Mar 2014
4.
Review the parameters in the Preview pane of the selected SOP.
5.
Using the Intensity Monitor group at the bottom of the Measurement pane:
a. Check the
Cell Parameters
and
Diluent Properties
.
If performing temperature gradient analyses, be sure you have selected one of the recom-
mended cells shown below. The analysis will not proceed with any other type.
Measurement
Cell Type
Range of Temperature
Size
Size cell (glass)
5.0 to 90.0
C
Size cell (Dispo)
5.0 to 50.0
C
Size cell (micro)
5.0 to 90.0
C
Zeta
Flow cell
5.0 to 90.0
C
Zeta cell (Dispo)
5.0 to 50.0
C