Group
Parameter
Description
General
(
continued
)
Manual
Temperature Setting
Enables you to specify the measurement temperature.
Select
Yes
; a
Temperature
field is inserted so that
you can enter a temperature. Enter a value between 5
and 90
C. If you specify a temperature less than 10
C
,
you should use nitrogen air to prevent
condensation on the cell surface.
Use this option when
using different temperatures for your analyses.
No
uses the temperature currently set in the intensity
monitor. Use this option when using the same
temperature for all analyses.
Gradient
allows you to analyze the same sample at
different temperatures; a
Gradient Temperature
field is added. Select the ellipsis at the right of the
field to display the Gradient Temperature Input
dialog.
Gradient Temperature Table
displays conditions
specified in the Input Gradient panel.
Select how the temperature is to be sorted;
Ascending
or
Descending
.
Use
Clear
to clear the table of its entries.
NanoPlus Operator’s Manual
SOP Designer
Mar 2014
3-11
Содержание NanoPlus
Страница 1: ...NanoPlus Zeta Potential Nano Particle Analyzer Operator s Manual V5 22 P09 42801 01 March 2014...
Страница 80: ......
Страница 114: ......
Страница 176: ...Consumables NanoPlus Operator s Manual A 30 Mar 2014 Tubing...
Страница 188: ......
Страница 192: ......
Страница 203: ...NanoPlus Operator s Manual AutoCorrelation ACF Function Mar 2014 D 11 Normal Log ACF ACF Noisy Low Sample Concentration...
Страница 213: ...NanoPlus Operator s Manual Temperature Gradient Plot Mar 2014 D 21 With Temperature Marker...
Страница 232: ......
Страница 246: ......
Страница 252: ......