HPVA Series Operator’s Manual
Adjusting the Metering Valves
Mar 2011
6-11
Adjusting the Metering Valves
The metering valves control the pressure of the gas flow into the unit.
It is very important that they are set correctly to ensure proper dosing
without overshooting pressure points.
There are three metering valves inside the instrument; one for each gas line. When you change gases or
need to adjust the flow of a gas follow these procedures.
1. Remove the rear panel from the instrument.
2. The metering valves are located on the lower right side of the instrument.
The valves are, from let to right, Gas 1, Gas 2, and Gas 3 (Helium).
Fully close the metering valve for the appropriate gas line.
3. Select
Manual Control
from the HPVA main menu.