System Configuration Window
HPVA Series Operator’s Manual
5-12
Mar 2011
Max Equil Time
The maximum time allowed for equilibration during the free
space analysis.
Temperature Tolerance
The temperature tolerance allowed for the ambient and analysis
temperatures during the free space analysis ( + this value).
Evac Expand
Evac Time
The amount of time the system will evacuate the sample tube prior
to the free space analysis.
Expand Time
The amount of time valve 1 is open
ed to allow the gas in the
manifold to expand into the sample tube during free space
analysis.