System Configuration Window
HPVA Series Operator’s Manual
5-6
Mar 2011
Instrument Options Tab
Servo Configured
If the instrument is configured with servo valves, this option must
be selected. If the instrument is configured with needle valves,
this option must be cleared.
Digital Output Tasks
Vacuum
Vent
The NI MAX task associated with the vacuum servo valve.
The NI MAX task associated with the vent servo valve.
Multiplier
Vacuum
Vent
Increase this value to increase the speed of the vacuum servo
valve; decrease it to increase the servo precision.
Increase this value to increase the speed of the vent servo valve;
decrease it to increase the servo precision.
Constants
These are fixed values used for the servo algorithm, they cannot
be changed by a user.
Delay After Pulse
Enter the time between each pulse made when using the servo
valves.
Pulse Minimum
Enter the minimum amount of time the servo valve will remain
open during a pulse
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...