Configuring the Hardware
HPVA Series Operator’s Manual
5-2
Mar 2011
Configuring the Hardware
When you select
Configure Hardware
on the Main Menu, the System Configuration window is
displayed.
System Configuration Window
The System Configuration window contains 8 tabs of parameters that you can edit prior to performing
an experiment. The tabs and their associated dialogs are described below.
Instrument Tab
Serial Number
Enter the instrument’s serial number. If a number is not entered
or a wrong number is entered, the software will be unable to
find the instrument’s calibration data.
Pressure Limit
Enter the highest pressure the manifold will be allowed to
reach – for a HPVA-100, the maximum should be set no higher
than 103 bar and for a HPVA-200, the maximum should be set
no higher than 202 bar.