Running the Experiment
Mar 2011
2-21
2. Choose one of the following:
• To enter a new set of experiment parameters, proceed to step 3.
• To use an existing set of experiment parameters:
a. Click
Recall Template
to display the Select Experiment Template File to Load window.
b. Select the template name, then click
Load.
c. Enter an experiment ID in the
Expt ID
field.
c. Make any desired changes.
d. Proceed to Step 10.
3. Enter an ID for the experiment.
4. Enter an Operator ID for the experiment.
5. Under
Use
Sample Ports
, click 1. The window shows ports 1 through 4; however, the models
covered in this manual, the HPVA-100 and HPVA-200, have only one sample port.
6. Enter the weight of the sample that was calculated when preparing the sample.
7. Enter the sample name and, optionally, the lot number and notes.
8. Enter Low and High Pressure Factors if desired.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...