Purging the System
HPVA Series Operator’s Manual
2-18
Mar 2011
14. When the system reaches full vacuum, open the isolation valve on the sample holder by turning it
completely counterclockwise.
Turn isolation valve
completely counterclockwise
15. Plug the temperature probe cable into the connector on the instrument.
16. Insert the temperature probe into the temperature control vessel.
Temperature
Probe
Connector
17. Add the appropriate liquid to the temperature control vessel until the level reaches about ½ inches
from the top.
18. Double-click the Bath icon and set the circulating bath controller to the desired temperature.