HPVA Series Operator’s Manual
Purging the System
Mar 2011
2-17
6. Open valve 8 and pressurize the instrument until the
Sample
pressure displayed on the screen
reads 15 bar.
24.00
31.00
15.000
15.000
7. Close valve 8.
8. Open valve 2.
9. Adjust the
Vent
control knob on the front panel of the instrument to allow pressure to vent.
10. When the pressure reaches 1 bar or less, close valve 2
11. Repeat steps 2 through 9 three times, then proceed to step 11.
12. Open valve 4 until the pressure reaches about 0.5 bar, then close valve 4.
13. Perform the unrestricted evacuation by opening valve 5.