Degassing the Sample
HPVA Series Operator’s Manual
2-10
Mar 2011
8. Slowly open the isolation valve on the sample holder by turning it completely counterclockwise.
Turn isolation valve
counterclockwise
9. Monitor the external vacuum gauge to ensure that vacuum is reached.
10. Set the temperature to the desired level by holding the left
FURNACE
button and the
up
or
down
button as needed. Allow the sample to degas as needed for the material to be analyzed.
11. To end the degas procedure, close the isolation valve by turning it completely clockwise.
Turn isolation valve
completely clockwise