Turning the Analyzer On and Off
HPVA Series Operator’s Manual
1-8
Mar 2011
Turning the Analyzer On and Off
To turn on the analyzer:
1. Place the power switch on the rear panel of the analyzer in the ON ( | ) position. The alarm
indicator will light indicating that power is applied to the instrument.
2. Wait about 20 seconds, then press the
ALARM RESET
button on the back panel of the analyzer.
3. The alarm indicator will turn off.
4. Wait 30 to 60 minutes to allow the manifold temperature to stabilize.
To turn off the analyzer:
1. Ensure that the analyzer is vented to atmosphere.
2. Place the power switch on the rear panel of the analyzer in the OFF (O) position.
Air
Connects the nitrogen or compressed air supply for the
pneumatic valves to the instrument.
Gas1
Connects an analysis gas to the instrument.
Gas2
Connects another analysis gas to the instrument.
Gas3
Connects helium (backfill gas) to the instrument.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...