HPVA Series Operator’s Manual
Controls and Connectors
Mar 2011
1-5
Vacuum Valve Control
Enables you to control the flow rate through vacuum
valve 4.
Vent Valve Control
Enables you to control the flow rate through the vent
valve.
Valve Switches
Enable you to open and close valves. You can also open
and close valves 1 through 8 through the software, which
is the recommended method of manual control.
Valve 9 can be controlled by the switch only.
System Schematic
A schematic of the manifold, showing valve status. A
valve indicator turns on when the valve is open and turns
off when a valve is closed.
When the HPVA is performing an experiment or being operated in
manual control mode, the valve switches on the front panel should be
placed and remain in the closed (down) position.
If a switch is placed in the opened position during a software
operation, it will override the software and may damage the
instrument.
Содержание HPVA Series
Страница 1: ...HPVA Series High Pressure Volumetric Analyzer Operator s Manual V4 0 P02 42801 01 March 2011...
Страница 4: ......
Страница 8: ...Table of Contents HPVA Series Operator s Manual iv Mar 2011...
Страница 20: ...Specifications HPVA Series Operator s Manual 1 12 Mar 2011...
Страница 52: ...Definitions of Fields in the System Windows HPVA Series Operator s Manual 2 32 Mar 2011...
Страница 66: ...Displaying Experiment Data HPVA Series Operator s Manual 3 14 Mar 2011...
Страница 70: ...Setting the Sample Bath Temperature HPVA Series Operator s Manual 4 4 Mar 2011...
Страница 98: ...Ordering Informatioin HPVA Series Operator s Manual 7 2 Mar 2011...
Страница 120: ...Analyzing the Results HPVA Series Operator s Manual B 18 Mar 2011...
Страница 124: ...Index HPVA Series Operator s Manual Index 4 Mar 2011...