background image

-2-

GENERAL SAFETY RULES

Please read all instructions. Failure to follow all instructions listed below may result in injury. If the
equipment is used in a manner other than as specified in these operating instructions, the
protection provided by the equipment may be impaired. Always wear eye protection.

DO NOT OPERATE

unit from a DC power source.

Do not abuse the power cords. Never carry the instrument by the cord or attempt to unplug the
instrument using the cord. Always operate the instrument with the standard installed cord.
Changing or using a damaged cord can increase the risk of electrical shock. The cord should be
checked periodically for any damage.

Do not position the instrument such that it would be difficult to operate the disconnect device
(plug) on the end of the power cord.

The outside housing should remain intact and solid. Any damage, chipping, or separating exposing
internal wires is a hazard. Instruments should not be used in this condition. The outside housing
should be periodically checked for damage.

Duty Cycle:

The 300 series are designed for a 50% duty cycle, or approximately two minutes on

and two minutes off. Continuous operation may cause overheating and damage the Contour Probe.

Operating Environment:

Temperature: 32° to 104°F (0° to 40°C). Relative humidity: 10% to 95%,

non-condensing

Shipping and Storage Environment:

Temperature: 40° to 140°F (4.44° to 60°C). Relative

humidity: 5% to 95%. Vibration and shock: As encountered in normal shipping and handling with 
no degradation.

General Cleaning

The outside surface of the instrument can be periodically wiped clean with a clean cloth and a
mild general purpose cleaner. Avoid using cleaners such as lacquer thinner, or mineral spirits that
could damage the outside housing.

Never attempt field service. All 300 series Contour
Probes should be returned to the factory for repairs.

OPERATION:

Connect the instrument plug into a

Grounded power outlet of proper voltage. Place the
Contour Probe legs upon the work surface with the
suspected defect at right angles to the legs (Good
contact will produce the best results). Push the test
switch to energize the instrument. Lightly dust or
float dry magnetic inspection powder over the area
being inspected. Defect indications will be revealed
in a direction shown on figure 2. Turn the Probe 90
degrees from the first test and repeat the process.
This method may also be utilized when applying a
wet medium. Check all procedures and standards for
further details involving inspection specifics.

FIG. 2

Отзывы: