background image

Diagnostic Test Plug

!

Do not throw away this plug.

!

!

The  Diagnostic  Test  Plug  should  only  be  used  by  a  technical  representative  from 

Parker  in  order  to  diagnose  compressor  issues.  Please  keep  this  plug  with  the 

generator’s documentation or accessories.

!

!

!

Description

:  The purpose of this test is to isolate and verify that the generator’s compressor reaches 

max  pressure  (specified  in  the  Table  in  Step  5)

  within  2  minutes.  If  the  Pressure  is  OK  then  the 

pressure problem is downstream of the compressor. 

!

Diagnostic Procedure

!

1. Power Down generator and allow enough time for the generator to depressurize, verify by

observing the generator’s pressure gauge located on the front panel. 

!

Note: It can take up to an hour if fully pressurized.

!

!

2. On the back panel, locate the air outlet connection (not labeled) that

connects to black separator assembly.

!

!

3. While pushing down on the connection’s collar, pull to remove the

3/8” tubing from this location.

!

!

!

!

!

4. Insert The Diagnostic Plug into this connection.

!

!

!

!

!

5. Power ON generator, and time how long the generator takes to come up to max pressure and

shuts off the pressure switch. Refer to chart below for times.  If test passes, then the problem is
downstream of the compressor. If test fails a problem resides with compressor.  

!

a. Check for leaks. For troubleshooting refer to Manual or call Technical Services for assistance

1.800.343.4048

!

!

 TI-BT_PLUG  rev.C

Model

Pressure Switch OFF 

Setting

Time to Shut OFF

LCMS-5000

0.0 psig - 145psig

< 2 min 20 secs

NITROFLOW

60

NITROFLOWTG1

0.0 psig - 125psig

< 2 min.

NITROFLOWTG2

0.0 psig - 145psig

< 2 min 20 secs

Collar

Отзывы: