background image

ULTRASONIC SCALER

h

 D660

MD   Rx ONLY

This manual, in whole or in part, should not be considered a substitute for formal 
training in ultrasonic scaling. Appropriate dental professional education is strongly 
recommended prior to using this device clinically.

A01156revA0722

English

Instructions for Use

Parkell Technical Support Service

(800) 243-7446
M-F from 8:30 AM to 5:00 PM EST

Содержание TurboSensor +

Страница 1: ...idered a substitute for formal training in ultrasonic scaling Appropriate dental professional education is strongly recommended prior to using this device clinically A01156revA0722 English Instructions for Use Parkell Technical Support Service 800 243 7446 M F from 8 30 AM to 5 00 PM EST ...

Страница 2: ... of Power Cord 110V 12 6 380cm 220 230V 7 6 230cm Handpiece Operating Frequency 30 KHz hD722 Protection Against Electric Shock Class 1 Type B applied part Protection Against Ingress of Liquids Foot Switch Scaler IPX1 drip proof Power Supply IPX0 Ordinary Mode of Operation of Equipment Intermittent 10 minutes on 5 minutes off Operating Conditions 15 30 C 20 70 RH non condensing Transport and Storag...

Страница 3: ... handpiece cable and the handpiece are clean and dry before assembling them for use As with all electrical devices the unit should not be immersed in fluids and fluids should not be allowed to enter the unit through front or rear panel slots Do not reach for the device if it has fallen into liquid until power is disconnected and do not use the device Return it to Parkell for servicing Modifying de...

Страница 4: ...foot from pedal to evaluate deposit removal Exercise caution near ceramic or resin restorations as they can fracture if stressed Inserts Magnetostrictive Inserts for the TurboSensor scaler are available separately and are not included with the basic scaler unit The unit is designed to work with all Parkell and other market leading 30KHz inserts Brand new inserts may fit very tightly in the handpie...

Страница 5: ...o enter Turbo mode A yellow indicator light will illuminate on the front panel to confirm the scaler is operating at Turbo power Scaling power to a point midway between the current setting and the scaler s maximum power Engaging the Turbo feature at a low power setting causes a significant boost in power Engaging the Turbo feature at high power will produce very little change Since water flow is n...

Страница 6: ...ace You may gouge the tooth Pause occasionally during scaling by removing foot from pedal to evaluate deposit removal Exercise caution near ceramic restorations as they can discolor or fracture if stressed Common Operator Errors that May Result in Poor Clinical Performance Failure to keep insert tip parallel to the long axis of the tooth Excessive use of hand pressure Failure to use the lowest eff...

Страница 7: ...logged Inspect and clear blockage or replace insert Water filter clogged Replace No water supply connected Correct Insert does not go into handpiece Insert stack is bent Straighten plates carefully by hand and reinsert O ring is too tight Lubricate O ring with water and use twisting motion to seat insert Detachable 30K Handpiece hD722 can only accept 30K insert Make sure that it is not a 25KHz ins...

Страница 8: ...essories with the unit and include an explanation of the problem Pack in the original box add plenty of cushioning material and over box the unit during shipping Transit should maintain a dry temperature of 0 F 110 F You will be contacted for your approval of the repair along with any associated costs prior to any work The unit will be repaired and returned to you Outside the US repairs must be ma...

Отзывы: