PHYSIO-PORT
User Manual Version 2017-11-12
46
10.
Cleaning and Maintenance
10.1
Cleaning and disinfection of device surface
Switch off the PHYSIO-PORT Recorder.
Wipe the device down with a soft, lint-free cloth for cleaning. Liquids must not penetrate the device.
Spray disinfection has proved successful. Incidin® Foam or equivalent disinfectants that are used in
practices or hospitals are suitable (Please respect the information of the manufacturer especially
regarding the exposure time).
10.2
Cleaning and disinfection of cuffs
Light stains can be removed by using a moist cloth.
For heavier stains, rinse the cuff with soapsuds or disinfected cleaning agent (not in the washing
machine!). No fluid should enter the cuff inflation section or the connecting tube.
Spray disinfection has proved successful after the usage of the cuff. Incidin® Foam or equivalent
disinfectants that are used in practices or hospitals are suitable (Please respect the information of the
manufacturer especially regarding the exposure time).
After you have cleaned the cuff, you need to rinse it thoroughly with water and let it dry at room
temperature for approx. 15 hours.
CAUTION
• Danger of electric shock •
Always disconnect the device from your PC before
cleaning it
.
• Damage to the device •
Do NOT use any disinfection agents that contain
phenol and peroxide for disinfecting the surface
of the device!
If any fluid has entered the device, then it should
only be used again after having been checked
and approved by the Service Department.