Maintenance & Repair
Repair work may only be carried out by personnel who have been authorized by the
manufacturer.
In the event of a malfunction or deficiency, the device can only be repaired by a qualified
technician approved by panno-med. In such an event, please mark clearly that the device is
OUT OF SERVICE. Repairs performed by unqualified personnel may result in serious
damage or injury.
Regular maintenance check
It is recommended to perform a regular, annual inspection of the device by trained
personnel that are authorized by the manufacturer.
This inspection must include at least the following aspects:
1.
Comprehensive visual and functional test
a.
Visual inspection
➢
All welding seams for cracks
➢
Surface damages
➢
Damages to the table top
➢
All cables (bruises, abrasions, exposed wiring)
➢
Visual inspection for leak-tightness of hydraulics
➢
Hinges and swivels
➢
Safety locks (circlip, screws, star lock) on all swivel joints, hinges, and motor
suspensions
➢
Visual inspection for completeness and damage of the plastic bushings in the
actuator mountings
➢
Visual inspection of all gas springs for leak-tightness
b.
Functional test
➢
Height adjustment
➢
Tilt adjustment
➢
Operation of castors
➢
Push handles
➢
Moveable components
➢
Chassis and suspension
2.
Function of the operational elements and check of the table control for damage
3.
Battery and control unit are examined for charging activity
4.
All bolt & screw connections inspected for tightness
5.
Electrical test
6.
Safety check
Содержание VET LIFT 7H9.6P1
Страница 14: ...Instruction Manual VET LIFT Page 14 of 27 Version 2 02 ...
Страница 15: ...Instruction Manual VET LIFT Page 15 of 27 Version 2 02 ...
Страница 16: ...Instruction Manual VET LIFT Page 16 of 27 Version 2 02 ...
Страница 21: ...Instruction Manual VET LIFT Page 21 of 27 Version 2 02 ...
Страница 22: ...Instruction Manual VET LIFT Page 22 of 27 Version 2 02 ...