Chapter 2. Initial Setup
58
Sentinel™ User’s Manual
2.6.1.3
Errors
To set limits for possible errors, see
and consider the following:
Note:
In the event that one of the following errors exceeds its limit, a corresponding error notice, described in
Chapter 4,
Troubleshooting
, will be displayed by
PanaView
and
Modbus.
Also, the alarm relay will change
state and the 4-20mA output and the Frequency output will be able to be programmed to identify the
presence of an error.
•
Soundspeed:
The soundspeed % Theory is set to cover the range at which the gas soundspeed may vary as a
function of its composition, pressure and temperature. Setting the number too low may cause the meter to get
into an error mode and stop working. Setting the number too high may cause the meter to fail to detect
instrument malfunctions.
•
Signal Strength Limits:
The meter can operate in a very wide range of signal strength. The low limit should be
set to allow the Signal to Noise Ratio to be large enough to ensure that the meter will maintain the specified
accuracy. The upper limit needs to be set to ensure that the meter will not be overloaded and the signal be
undetected.
•
Amplitude Limits:
The steady state amplitude is 100%. In case of a sudden change in the system gain, it may
take time for the AGC to stabilize the amplitude to 100%. An error will be indicated if the amplitude exceeds the
limits.
•
Velocity:
The velocity limits should be set higher than the maximum possible flow.
•
Acceleration:
The meter is testing for a change in velocity differential between each set of two consecutive
velocity calculations. In some applications a sharp change in velocity is expected. In that case the meter
should be programmed with a higher value than the default, which is 1.5 m/s.
Содержание Sentinel
Страница 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents x Sentinel User s Manual...
Страница 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Страница 55: ...Sentinel User s Manual 55 Chapter 2 Initial Setup 2 6 1 1 General cont Figure 13 General Tab Display...
Страница 57: ...Sentinel User s Manual 57 Chapter 2 Initial Setup 2 6 1 2 Fluid cont Figure 14 Fluid Tab Display...
Страница 59: ...Sentinel User s Manual 59 Chapter 2 Initial Setup 2 6 1 3 Errors cont Figure 15 Errors Tab Display...
Страница 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Страница 90: ...Chapter 3 Operations Sentinel User s Manual...
Страница 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Страница 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Страница 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Страница 132: ...Appendix B Data Records 132 Sentinel User s Manual B 2 Diagnostic Parameters Date Description of Service Performed By...
Страница 135: ...Sentinel User s Manual 135 Appendix C Brazilian INMETRO Approval Appendix C Brazilian INMETRO Approval...
Страница 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Страница 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Страница 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Страница 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Страница 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Страница 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Страница 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Страница 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Страница 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Страница 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Страница 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Страница 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Страница 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Страница 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Страница 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Страница 158: ...Index 158 Sentinel User s Manual...
Страница 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Страница 162: ......
Страница 164: ......
Страница 165: ......