Chapter 2. Initial Setup
54
Sentinel™ User’s Manual
2.6.1
Channel Tabs
Use this section to program the meter parameters, which requires the user to login before any parameters can be
changed.
2.6.1.1
General
shows a typical
General
tab display:
•
Path Length (P):
This is the acoustics path length. The value is measured on the assembled meter.
•
Axial Distance (L):
The value of the axial distance is measured on the assembled meter or derived from Quality
Control Report measurement.
•
Pipe Diameter:
This is the inside diameter of the meter. This parameter is used by the meter as a factor in
volumetric calculation. The value is measured on the assembled meter or derived from Quality Control Report
measurement.
•
Transducer Frequency:
Select from a drop down menu the specified frequency for a pair of transducers.
•
Fine/Coarse Switch Point:
This is the number of points corresponding to the delta T at which the velocity
calculation will switch from Mode 2 to Mode 3.
The formula for converting from velocity to count number is:
Where,V = velocity (meters/second)
SOS = speed of sound (meters/second)
L = axial length (meters)
F = programmed transducer frequency (hertz)
Note:
The meter will be shipped with the correct count number to insure that the meter operates correctly.
•
# in Avg:
This is the number of parameters averaged together before being reported on average outputs. For
the critical parameters, like velocity and volumetric, it is the number of averaged measurements. This number
is carefully selected by the manufacturer to meet the meter specification as well as optimize the meter
performance.
•
Time delay (Tw):
The delay time is a number which includes various delays in the transducers, electronics and
cables. The manufacturer determines the exact Tw number during the zero flow calibration procedure.
•
Zero cutoff:
The value below which the velocity reading is forced to zero.
Counts
V
32LF
SOS SOS
¥
-------------------
¥
=
Содержание Sentinel
Страница 1: ...panametrics com Flow 910 246D August 2021 Sentinel User s Manual...
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 10: ...Contents x Sentinel User s Manual...
Страница 36: ...Chapter 1 Installation 26 Sentinel User s Manual...
Страница 55: ...Sentinel User s Manual 55 Chapter 2 Initial Setup 2 6 1 1 General cont Figure 13 General Tab Display...
Страница 57: ...Sentinel User s Manual 57 Chapter 2 Initial Setup 2 6 1 2 Fluid cont Figure 14 Fluid Tab Display...
Страница 59: ...Sentinel User s Manual 59 Chapter 2 Initial Setup 2 6 1 3 Errors cont Figure 15 Errors Tab Display...
Страница 82: ...Chapter 3 Operation 82 Sentinel User s Manual...
Страница 90: ...Chapter 3 Operations Sentinel User s Manual...
Страница 100: ...Chapter 5 Diagnostics 100 Sentinel User s Manual...
Страница 114: ...Chapter 6 Transducer Replacement 114 Sentinel User s Manual...
Страница 130: ...Appendix A CE Mark Compliance 130 Sentinel User s Manual...
Страница 132: ...Appendix B Data Records 132 Sentinel User s Manual B 2 Diagnostic Parameters Date Description of Service Performed By...
Страница 135: ...Sentinel User s Manual 135 Appendix C Brazilian INMETRO Approval Appendix C Brazilian INMETRO Approval...
Страница 136: ...Appendix C Brazilian INMETRO Approval 136 Sentinel User s Manual...
Страница 137: ...Sentinel User s Manual 137 Appendix C Brazilian INMETRO Approval...
Страница 138: ...Appendix C Brazilian INMETRO Approval 138 Sentinel User s Manual...
Страница 139: ...Sentinel User s Manual 139 Appendix C Brazilian INMETRO Approval...
Страница 140: ...Appendix C Brazilian INMETRO Approval 140 Sentinel User s Manual...
Страница 141: ...Sentinel User s Manual 141 Appendix C Brazilian INMETRO Approval...
Страница 142: ...Appendix C Brazilian INMETRO Approval 142 Sentinel User s Manual...
Страница 143: ...Sentinel User s Manual 143 Appendix C Brazilian INMETRO Approval...
Страница 144: ...Appendix C Brazilian INMETRO Approval 144 Sentinel User s Manual...
Страница 145: ...Sentinel User s Manual 145 Appendix C Brazilian INMETRO Approval...
Страница 146: ...Appendix C Brazilian INMETRO Approval 146 Sentinel User s Manual...
Страница 148: ...Appendix D NMI Nederlands Meetinstituut Approval 148 Sentinel User s Manual...
Страница 150: ...Appendix E Romanian Bureau of Legal Metrology Approval 150 Sentinel User s Manual...
Страница 151: ...Sentinel User s Manual 151 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 152: ...Appendix E Romanian Bureau of Legal Metrology Approval 152 Sentinel User s Manual...
Страница 153: ...Sentinel User s Manual 153 Appendix E Romanian Bureau of Legal Metrology Approval...
Страница 154: ...Appendix E Romanian Bureau of Legal Metrology Approval 154 Sentinel User s Manual...
Страница 158: ...Index 158 Sentinel User s Manual...
Страница 160: ...Warranty clx Sentinel User s Manual no content intended for this page...
Страница 162: ......
Страница 164: ......
Страница 165: ......