5-2
Clearing Data
The TOTAL Sub-Menu
The
TOTAL
sub-menu permits the user to reset volumetric and energy
totals for channel 1, channel 2 or both channels to zero. In addition,
the stopwatch totalizer is reset. Press the appropriate side of the
[SCREEN]
key to activate the desired pane of the display screen, and
refer to Figure A-7 on page A-7 of Appendix A, Menu Maps, while
completing the following steps:
Ch1 Channel Label site file
É
The default power up screen is in
BIG
mode. Access the Clear
Menu by pressing the
[CLR]
key.
Velocity Ft/s
6.95
(error codes appear here)
VEL
VOLUM +TOTL
-TOTL
OPEN Start
Enter the assigned password and
press
[ENT]
. If the security
feature has been deactivated, this
prompt does not appear. See
Chapter 1, Programming Site
Data, for details.
ENTER PASSWORD
Password
CLEAR Start
Press
[F1]
to select the
TOTAL
option.
CLEAR
previous selection appears here
TOTAL
SITE
LOG
CLEAR TOTALS
Press
[F1]-[F3]
to select the
channel(s) to be cleared.
CLEAR
Clear Total
Total to Clear
previous selection appears here
CH1
CH2
BOTH
CLEAR TOTALS
Press
[F1]
to abort the procedure
or press
[F2]
to clear the selected
totals and reset the stopwatch
totalizer. In either case, the initial
Clear Menu prompt returns.
Total to Clear
current selection appears here
CLEAR CHx TOTALS
previous selection appears here
NO
YES
Содержание GF868
Страница 1: ...Model GF868 Programming Manual Two Channel...
Страница 6: ...Chapter 1...
Страница 71: ...Chapter 2...
Страница 92: ...Chapter 3...
Страница 113: ...Chapter 4...
Страница 131: ...Chapter 5...
Страница 132: ...Clearing Data Introduction 5 1 The TOTAL Sub Menu 5 2 The SITE Sub Menu 5 3 The LOG Sub Menu 5 4...
Страница 138: ...12 1 99 5 6 Clearing Data...
Страница 139: ...Chapter 6...
Страница 150: ...Appendix A...
Страница 159: ...Appendix B...
Страница 160: ...Data Records Option Cards Installed B 1 Initial Setup Data B 2...