DigitalFlow™ XMT868i Service Manual
25
Appendix B. Calibration and Testing with PanaView
B.4.3 RTD Inputs
RTD inputs may be added to the Model XMT868i by installing an appropriate option card in Slot 1. This option card
contains two or four RTD inputs, which are designated as A, B, C and D. Both the set and slope temperature values for
each input must be calibrated.
Calibration of the RTD inputs requires the use of a calibrated RTD source.
Prepare for the calibration procedure by connecting an independent calibrated RTD source to the desired input of
the option card. See Figure 4 on page 12 to identify the RTD (+) and COM(-) pins on terminal block J2.
1.
In the center pane of the
Calibration/Test
window, double-click on Slot 1.
2. Double-click on the desired input to open the
RTD Calibration
menu.
IMPORTANT:
The procedure for calibrating
each input is identical.
However, be sure to reconnect the temperature
reference on terminal block
J2
.
for the correct pin numbers.
Proceed to the corresponding sub-section for specific instructions on each of the options shown above.
B.4.3a Set Point Option
1.
If you selected the
Set Pt.
option in the
RTD Input
menu, PanaView asks for the
Low Reference Temp.
Enter the low
reference value in the right pane, and click [Next Item].
2. Double-click on
Store
to accept the low reference value or on
Abort
to cancel the entry. In either case, the RTD
Input prompt reappears.
The XMT displays an
accepted
message if the value is good or an
out of limits, rejected
message if the value is bad. If
the value is rejected, try again.
B.4.3b Slope Point Option
1.
If you selected the
Slope Pt.
option in the
RTD Input
menu, PanaView asks for the
High Reference Temp.
Enter the
high reference value in the right pane, and click [Next Item].
2. Double-click on
Store
to accept the low reference value or on
Abort
to cancel the entry. In either case, the RTD
Input prompt reappears.
The XMT displays an
accepted
message if the value is good or an
out of limits, rejected
message if the value is bad. If
the value is rejected, try again.
Click [Exit Page] to return to the Slot 1 window and calibrate additional inputs/outputs, or click [Exit Page] again to
close the
Calibration/Test
menu.
Содержание DigitalFlow XMT868i
Страница 2: ......
Страница 4: ...ii no content intended for this page...
Страница 24: ...Chapter 2 Error Codes 4 DigitalFlow XMT868i Service Manual...
Страница 34: ...Chapter 3 Diagnostics 10 DigitalFlow XMT868i Service Manual...
Страница 44: ...Chapter 4 Parts Replacement 10 DigitalFlow XMT868i Service Manual...
Страница 48: ...Chapter 4 Parts Replacement DigitalFlow XMT868i Service Manual...
Страница 52: ...Appendix A Service Record 18 DigitalFlow XMT868i Service Manual...
Страница 66: ...Appendix B Calibration and Testing with PanaView 32 DigitalFlow XMT868i Service Manual...
Страница 72: ...Warranty 38 DigitalFlow XMT868i Service Manual no content intended for this page...
Страница 73: ......